III. Initial Installation
14
Dakewe (Shenzhen) Medical Equipment Co., Ltd.
www.dakewemedical.com
Before connecting the instrument to mains for the first time, be sure to keep the
power switch at
“〇”
.
Daily startup of the device: Select the ON/OFF button on the right side of the
device (Figure 4-17), and briefly press the ON/OFF button.
While operating, please do not turn off the instrument.
Please follow the steps provided above to completely shut down the instrument.
Turning off the instrument in other ways may cause serious damage to the
instrument hardware and result in data loss.
3. Turning off the Instrument
The following three steps are required to completely turn off the Dakewe Slide Printer:
(1) Long press the ON/OFF button in the front of the instrument for 3 seconds so that the software will
pop up the account logout prompt to enter the shutdown process.
(2) After the touch screen turns off, the shutdown process is complete.
(3) Instrument outage: Select the main power switch on the back of the instrument to keep the switch
at the “
〇
” end (this step is not necessary during the use of the instrument).