III. Initial Installation
Dakewe (Shenzhen) Medical Equipment Co., Ltd.
9
www.dakewemedical.com
Figure3- 3 S100-Front view of the Dakewe Slide Printer
(2) Requirements for Installation
The device requires an installation area of about (L*W*H) 600*400*800 mm.
The device power supply range is 100
–
240VAC and the power cord should be a dedicate cable
that conforms to the local power supply.
The device should be installed at a place where the main power switch can be easily accessible at
any time. Never connect the device to a power strip. Instead, it must be connected to a grounded mains
socket-outlet.
The operator should ensure that electrostatic discharge (ESD) safety precautions are in place.
The device is for indoor use only. Do not use the device indoors where there is a risk of explosion.
Avoid collisions, direct sunlight, and frequent electricity fluctuation.
Considering the weight of the device, the workbench should have the capacity and stiffness to
carry the weight of the device.
The room temperature should be always maintained b10
°
C and +35
°
C; the
installation site must be well ventilated and free from any sources of ignition.
Figure3- 4Rear view of S200
1 Speaker
2 Vent 1
3 Vent 2
4 Air outlet
5 Vent 3
6 Laser air inlet
7 USB Ports / Ethernet Ports
8 Power switch