DAIHAN Scientific, Steam Sterilizers Digital Fuzzy-controlled Autoclaves .
20/53
②
MAIN - Set Temperature
-
If pressing once the MAIN button in the Set Operation Time Mode, it is switched to the
temperature setting mode.
-
The setting temperature is entered by turning the Jog-Dial Knob to clockwise or counter-
clockwise. If the setting value is entered to above 0
℃
, the operation standby state is
displayed while blinking the HEATING indication of LCD Display.
-
The HEATING indication of LCD Display is always displayed during the operation of
HEATING, and if the setting value is entered to above 0
℃
in the condition that the heating
is stopped, the HEATING indication indicates the standby status while blinking in long
interval.
-
The range of the setting temperature is adjustable in the unit of 1
℃
, and can be set up to
132
℃
.
-
You can change or save the setting by using the operation method of the Jog-Dial, and
Start/Stop of operation can be performed.
* SUB - Offset Temperature
-
If pressing the SUB button in the Set Temperature mode, it is switched to the Set
Temperature Offset mode.
-
Temperature compensation is the function that can add or reduce the constant value by
setting the compensation value in the case that user must experiment by matching the
thermometer which he/she is using as the experimental standard and the thermometer
indicated in this regulator.
-
The scope of the compensative value is -10.0
℃
~ +0.0
℃
, and it is possible to adjust at 0.1
℃
unit.
< Fig 3.3.3. Set Temperature Mode >
Temp SV : 0
℃
Temp PV : 2 7
℃
Heating LED (Red)
Set Temp
Present Temp
< Fig 3.3.4. Set offset value Mode>
Temp PV : 27.0
℃
offset : 0.0
℃
Heating LED (Red)
Present Temp
offset Temp