DAIHAN-Sci.com or
ALL
for
LAB.com
Shopping Mall
DAIHAN Scientific, Steam Sterilizers Digital Fuzzy-controlled Autoclaves .
11/53
Functional Block Diagram
Page 1: ...dation IQ OQ Service Patented No 10 0554098 Jog Shuttle Control System User Manual Version 2 2 1 written for MaXterileTM 47 60 80 100 MaXterileTM 60 Steam Sterilizers Digital Fuzzy controlled Autoclav...
Page 2: ...rol with monitoring of final assessment Category II Autoclave 1 Item Vessel of Autoclave 2 Model names MaXterile 47 MaXterile 60 MaXterile 80 3 Design Pressure 2 5bar 4 Operation Pressure 1 2 2 bar 5...
Page 3: ...product Autoclave MaXterile 100 to which this declaration relates is in conformity with the following standard or other normative document EN 61326 EN 61000 3 2 EN 61000 3 3 EN 61010 1 EN 61010 2 foll...
Page 4: ...ore Use 5 1 2 Product Introduction 6 1 3 Product Configuration 7 2 Installation 2 1 Product Installation 8 2 2 Product Specifications 10 3 Product Usage 3 1 Product Usage 12 3 2 Names of Each Part 15...
Page 5: ...well acquainted with the meaning Warning Failure to follow the instructions may result in injury Caution Failure to follow the instructions may result in product failure Proximity Caution Keep clear...
Page 6: ...ess steel material processed by a press is used in order to maintain the aesthetics of Chamber and the high level of hermeticity and the outside overheating of the equipment is protected against in th...
Page 7: ...igital Fuzzy controlled Autoclaves 1 Unit Wire Basket Set 2 ea Heater Cover 1 ea Drain Bucket 1 ea Including a copy of the Operation Manual Model MaXterileTM 47 Steam Sterilizers Digital Fuzzy control...
Page 8: ...e inform the forwarder and your contractor If there is no damage please go to next point 3 4 Let the unit stand for a few minutes so that the all the internal parts and components of the unit can be s...
Page 9: ...nsurance so that its customers can trust our products believe and use its products by preparing against the accident which can be occurred during using the products by self production Contents Insuran...
Page 10: ...Function Controller Display Fuzzy Control System Fully Automatic Door Closing System Lever type Sliding Door Safety Electronic Door Closing System Safety Circuit Over Temp Pressure Current protection...
Page 11: ...DAIHAN Sci com or ALL forLAB com Shopping Mall DAIHAN Scientific Steam Sterilizers Digital Fuzzy controlled Autoclaves 11 53 Functional Block Diagram...
Page 12: ...esigned for the following purposes For sterilizing glassware rubber items pipette tips instruments filter bodies as well as consumables waste in bags and to destroy infectious or septic materials For...
Page 13: ...stone a metal or paper goes insides it may cause a fire an electric shock or a break down If do pull out the mains plug from a socket and contact a manufacturer or a supplier Do not disassemble or re...
Page 14: ...s or instruments made from steel or iron Metering equipment unless their suitability for stream sterilization is specifically mentioned Do not open sterilizing chamber until full sterilization operati...
Page 15: ...ge Breaker Circuit Breaker Power Switch Power ON OFF Lever Door Safety Lock and easy open Lid Clamp Door Locking Control Panel LCD Display Pressure Gauge manometer Moving Wheel Transport Device Table...
Page 16: ...Names Description Door Fastener Door locking mechanism Safety Knob Prevent overheat Manual purge valve pressure pass Control Device Water Drain Valve Drain Control Device Condenser emission steam con...
Page 17: ...of TIMER HEAT ERROR MAIN Button Convert into mode of Timer Heating Printer Locking SUB Button Convert into mode of delay timer offset function Setting Operating Information OPEN Door open Jog Dial kno...
Page 18: ...turning the Jog Dial knob to clockwise or counter clockwise the setting values can be adjusted and if briefly pressing once the PUSH button the Start or the Stop operation is performed The Turning of...
Page 19: ...y setting the reservation time value with the Jog Dial knob the value is fixed on the Display and the reservation timer is activated After setting the reservation timer value you can set up the operat...
Page 20: ...blinking in long interval The range of the setting temperature is adjustable in the unit of 1 and can be set up to 132 You can change or save the setting by using the operation method of the Jog Dial...
Page 21: ...itched to the Set Temperature Mode MAIN Safety Locking If pressing the MAIN button in the Set Temperature mode it is switched to the Safety Locking Mode In the Locking Mode the Turning by the Jog Dial...
Page 22: ...he steam is rapidly and safely emitted through the emission steam condensation mechanism In LIQUID cycle mode valve is not completely opened at a time Firstly it waits for 5 minutes with valve closed...
Page 23: ...eck If the user wants to remove the remaining distilled water in the pressure vessel it can be removed by opening the Water Drain Valve Please check the device which will collect the drained distilled...
Page 24: ...nagement 4 1 Accessories Ordering Information Image Ordering Number Information DH WAC20047 Perforated Basket 270 x h250mm WAC20047 for 47Lit MaXterileTM 47 MaXterileTM 47R DH WAC20060 Perforated Bask...
Page 25: ...ay cause an electric shock or fire To clean the unit a neutral detergent and soft cloth is recommended Do not use coarse cloth strong chemicals or organic solutions Do not use water Benzene Thinner or...
Page 26: ...e of the device If the temperature does not rise at room temperature Check the temperature currently set Make sure that the power switch is in ON position Make sure that the Push button of the Jog Dia...
Page 27: ...t This instrument is warranted for 2 years from purchasing date A breakdown caused by customer s improper use or natural disaster is excepted from this warranty This certification is only for internat...