DAIHAN-Sci.com or
ALL
for
LAB.com
Shopping Mall
DAIHAN Scientific, Steam Sterilizers Digital Fuzzy-controlled Autoclaves .
19/53
-
If the timer operation is completed, the alarm will sound and heating will be stopped.
-
The timer setting mode can be operated by referring to the jog-shuttle operation method.
* SUB - Set Delay Time
-
If pressing the SUB button in the Set Operation Timer Mode, it is switched to the Set
Delay Time mode.
-
The Delay Timer sets the reservation time and has the function that sets so that the
operation of Heating and Timer are performed after that time is elapsed.
-
The proper use of the delay timer is to manipulate so that the reservation time is set for the
first time, and then the temperature and the operation timer is set and the set operation is
automatically operated after the reservation time is elapsed.
-
The reservation time value is displayed over blinking on the display, and if pressing the
PUSH button by setting the reservation time value with the Jog-Dial knob, the value is fixed
on the Display, and the reservation timer is activated.
-
After setting the reservation timer value, you can set-up the operation timer value by
pressing the MAIN button and then press Start button. And, you can set-up the temperature
by pressing the MAIN button again, and then press Start button. The Start button in the
condition that the reservation timer has been set-up, is the button for the operation of the
setting value after the reservation timer, not to immediately operate.
-
As described above, if continuously setting the operation, Heating and Timer are operated
after the set reservation time as user intend.
-
Since the delay timer is not set during the operation of Heating/Timer, the delay timer
should be set after stopping the operation of Heating/Timer.
-
You can change or save the setting by using the operation method of the Jog-Dial, and
Start/Stop of operation can be performed.
-
If pressing MAIN button, it is switched to the Set Operation Time Mode.
< Fig 3.3.2. Set Delay Time Mode>
DTim SV : 00:00
DTim PV : 00:00
Set Time
Remain Time
Timer LED (Green)