DAIHAN Scientific, Steam Sterilizers Digital Fuzzy-controlled Autoclaves .
13/53
You must wear the following protective items when opening the sterilizing chamber
and removing the sterilized articles;
-
Apron
-
Eye protectors
-
Heat insulated gloves
Watch for hot surfaces and hot steam!
After sterilizing liquids;
-
Danger from retardation of boiling.
-
For open vessels containing up to 1 liter: wait at least 2.5 hours
-
For open vessels containing up to 2 liters: wait at least 3 hours
-
Do not remove flasks or bags until the chamber temperature drops below
70
C.
Disconnect the power supply plug if the unit is not going to be used for an
extended period of time.
Step back when opening the sterilizer.
Do not put any objects insides.
When an object that is easy to be combustible is like a stone, a metal or paper
goes insides, it may cause a fire, an electric shock or a break down.
If do, pull out the mains plug from a socket and contact a manufacturer or a
supplier.
Do not disassemble or remodeling by yourself.
It may cause damage to the designed safety.
We do not have any duties for an accident, quality deterioration or a defect caused
by a customer.
All the mechanisms and electric/electronic devices in this machine are designed
and developed by the manufacturer. The devices could be disassembled and
serviced only by the engineers of manufacturer or by engineers approved by
manufacturer.
Please avoid ultrasonic lamps.
The unit could be cracked or faded.
Watch for hot steam and hot water.
Hot steam may escape when the lid is opened.
The lid handle should only be gripped at the experiment end.
Do not use outdoors avoid overheating or electrical faults. The use in the place
where the sun continuously shines and can get rained on it is prohibited. It may
cause overheating or electrical fault of the product.