DAIHAN-Sci.com or ALL for
LAB.com Shopping Mall
15/20
This key works in the same way as in General Weighing mode.
The position of the current weight is displayed with a graph in the same way as in General
Weighing mode.
(5) Setting a reference value of counting
It is possible to select one of 10, 20, 30, 40, 50, 60, 70, 80, 90, and 100 as a reference quantity.
When ENTER key is pushed,
‘REF.Set’ message flickers and
Setting mode appears.
Of Ld 10 -d 100, the latest set number
appears.
Push this key repeatedly until a proper
quantity appears.
Put the part or weight equal to the displayed
quantity on the platter and push ENTER key.
‘do.InG’ message appears first and then
‘SET OK’ message appears. The setting process
finishes and the scale automatically goes to
Counting mode.
At this time, the current quantity is displayed,
And counting process can be performed.
When ZERO key is pushed while setting,
the setting process is cancelled.
EXIT message appears and the scale returns to
Counting mode.
(6) Comparison weighing
Comparison weighing is used when a user sets a target value and compares the target value with a sample
weight.
A difference between a target value and the current weight value is shown on the Display part.
If the current weight value is lower than a target value, a negative value appears; if equal, ‘0’ appears; if
higher, a positive value appears.
Summary of Contents for Ba Series
Page 21: ......
Page 22: ......
Page 29: ...DAIHAN Sci com or ALL forLAB com Shopping Mall 7 21 3 3 1 AC 230V 50 60Hz 5 40 2 2 000 6 562 3...
Page 30: ...DAIHAN Sci com or ALL forLAB com Shopping Mall 8 21...
Page 31: ...DAIHAN Sci com or ALL forLAB com Shopping Mall 9 21...
Page 32: ...DAIHAN Sci com or ALL forLAB com Shopping Mall 10 21 10 31 80 40 50 1m 60dB...
Page 41: ...DAIHAN Sci com or ALL forLAB com Shopping Mall 19 21 4 4 2 080 008 3000 A S...