USER MANUAL
While using this device, I got a message saying that the temperature of this
device is high.
Turn off this device and let it cool down for approximately 1 hour prior to turning it back
on for use.
When the ambient temperature is high or this device is continuously used for a long
time, take a break from using this device. When you still get a message saying that the
temperature of this device is high even after taking a break, contact the service center
and the local person in charge.
I feel like the wrap is neither hot nor cold.
When the tubing or the wrap is bent, water will not flow.
When the ambient temperature is too high or too low, the cooling and heating may not
function properly.
When the peltier inside this device does not function properly, the cooling and heating
will not function properly. In this case contact the service center and the local person in
charge.
This device generates too much noise.
Check whether or not there is any foreign substance attached to the fan of the vent hole.
Check whether or not this device is placed on an unstable or tilted floor.
10-2. Error Code Occurrence
* If an error occurs during operation, the device displays the error code on the LCD screen.
Error Code
Symptom
Solution
Occurs when the internal
temperature of this device
increases
- Turn off and turn of this device.
- Turn off this device and let this device cool
down for approximately 1 hour prior to turning
it back on.
- When the error occurs when this device is
used after it is cooled down, contact the service
center and local person in charge.
Occurs when the
temperature of the heat
sink increases
Occurs when the water
container lacks water prior
to operating this device
- This device lacks water. Add more water to the
water container.
- Check whether or not this device is tilted.
Install this device on a flat floor.
Occurs when the water
container lacks water while
this device is in operation
- This device lacks water. Add more water to the
water container.
- Check whether or not the pipe of the wrap is
open. Check whether or not there is any water
leakage.
39