13
Note:
Filtration with thickness equivalency of aluminum.
- Using as method an x-ray tube tension of 70 kVp and 2.5 mmAl of semiconducting reductor.
-The method using the measurement of the first semireductor layer was in narrow beam conditions, with
the x-ray equipment operating with x-ray tube tension value of 70 kVp with anodic current of 7mA and
exposure time of 1.0 second.
X-Ray Spectro 70X
Mili Amper scale X Time
Set
Time
(seconds)
mA c/
Real
Time
Set x Real
KVp AVG
Set
Time
(seconds)
mA c/
Real
Time
Set x Real
KVp AVG
0,06
1,46
81,4
0,56
7,20
70,6
0,07
1,46
80,7
0,63
7,39
70,8
0,08
2,36
79,5
0,71
7,37
70,7
0,1
3,91
79,4
0,8
7,48
70,7
0,12
4,21
76,5
1
7,58
70,7
0,14
4,42
75,8
1,25
7,58
70,6
0,16
4,74
75,1
1,4
7,70
70,7
0,2
5,44
73,7
1,6
7,56
70,6
0,25
5,94
72,6
2
7,69
70,9
0,28
6,30
72,0
2,5
7,80
70,7
0,32
6 ,59
71,5
2,8
7,64
70,0
0,4
6,71
71,0
3,2
70,59
70,7
0,5
6,97
70,6
-----
-----
-----
Obs:
The specified field for time irradiation compliance is comprehended with in 0.06 and 3.2 seconds.
IMPORTANT:
Information referring to x-ray dosage, radiation leak can be found in the conformity
report attached to this manual.
- The serial number of the dome, beam limiter and radiation source is the same of the equipment,
and it is stated in the product’s label.
TECHNICAL SPECIFICATIONS
Summary of Contents for Spectro 70X Series
Page 1: ......
Page 15: ...15 Dimensions mm Model Coluna M vel TECHNICAL SPECIFICATIONS...
Page 16: ...16 Dimensions mm Model Parede TECHNICAL SPECIFICATIONS...
Page 17: ...17 Dimensions mm Model Pantogr fico Coluna M vel TECHNICAL SPECIFICATIONS...
Page 18: ...18 Dimensions mm Model Pantogr fico Piso TECHNICAL SPECIFICATIONS...
Page 19: ...19 Dimensions mm Model Pantogr fico Parede TECHNICAL SPECIFICATIONS...
Page 22: ...22 Content of accessible and non accessible demarcations 01 02 TECHNICAL SPECIFICATIONS...
Page 23: ...23 03 04 05 TECHNICAL SPECIFICATIONS Content of accessible and non accessible demarcations...
Page 42: ...42...
Page 43: ...43...
Page 44: ......