12
Distance protection against parasite radiation
The operator must use the safety device of the equipment (distance activation cable), remaining at
a minimum distance of 3 meters from the X-radiation beam during the charge application. The user
must wear the protection clothes (lead apron and lead collar) they are not provided with the product.
Lead apron and collar,
(They are not provided with
the product).
How to regulate arm tension
If the arm assembly/cylinder head loses its balance condition, a technician authorized by Alliage
shall be called in order to make the adjustment.
Radiation quality
Ingress of quality equivalence
Parametes for charge exposure 0,32s
FILTERS Tube crystal
1,26 mm Al
Insulating oil, column 17mm
0,64 mm Al
Window
0,32 mm Al
Additional filter (aluminum)
1,00 mm Al
TOTAL FILTRATION
3,22 mm Al
Tension
70 kVp ± 10%
Current
7 mA ± 1,5 mA
Power
0,49 kW
Product current time
2,24 mAs
Highest energy applied in intermittent
mode during 1 h
170 KJ
TECHNICAL SPECIFICATIONS
General features
Summary of Contents for Spectro 70X Series
Page 1: ......
Page 15: ...15 Dimensions mm Model Coluna M vel TECHNICAL SPECIFICATIONS...
Page 16: ...16 Dimensions mm Model Parede TECHNICAL SPECIFICATIONS...
Page 17: ...17 Dimensions mm Model Pantogr fico Coluna M vel TECHNICAL SPECIFICATIONS...
Page 18: ...18 Dimensions mm Model Pantogr fico Piso TECHNICAL SPECIFICATIONS...
Page 19: ...19 Dimensions mm Model Pantogr fico Parede TECHNICAL SPECIFICATIONS...
Page 22: ...22 Content of accessible and non accessible demarcations 01 02 TECHNICAL SPECIFICATIONS...
Page 23: ...23 03 04 05 TECHNICAL SPECIFICATIONS Content of accessible and non accessible demarcations...
Page 42: ...42...
Page 43: ...43...
Page 44: ......