DO NOT REPLACE LAMPS INDIVIDUALLY.  ALL LAMPS SHOULD BE REPLACED AT THE SAME TIME TO 
AVOID UNEVEN TREATMENTS OR BURNS.  PLEASE CONTACT THE DAAVLIN SERVICE DEPARTMENT 
PRIOR TO REPLACING YOUR LAMPS TO OBTAIN IMPORTANT INFORMATION ON HOW THE OUTPUT 
OF YOUR UNIT WILL BE AFFECTED BY THE CHANGE.   

Because  of  differences between  brands,  replace  lamps  with  the  same  brand  as  originally  installed.    A  Lamp 
Specification Guide is located below.  

 

Lamp Specification Guide  

 

 

 

 

 

Lamp Replacement 

1.  Consult the Lamp Specification Guide for replacement lamp specifications and contact the Daavlin 

service department for a lamp replacement quote.   

Note:

 The Daavlin service department can be reached toll free by dialing 1-800-322-8546 or via email 

at [email protected]

2.  Once you have received your replacement lamps unplug the machine. Then, using either a 5/16” hex 

driver/socket or a #2 Philips head screw driver, remove the screw from the bottom of each grid. 

3.  Lift each grid out of its holes in the bottom lamp plate, pull the bottom of the grid forward, then slide it 

out of the holes in the top of the unit and set it aside.   

4.  Grasp  the  lamp  to  be  removed  with  both  hands  and  press  down  until  it  clears  the  top  socket,  then 

remove the lamp.  Reverse the process to install lamps and grid. 

5.  Reset the lamp hours to zero.   

 

Resetting Your Lamp Hours  

Whenever the lamps are changed in your device, it is important to reset the lamp age to zero so the new 
lamp’s operation hours can be tracked. Please contact the Daavlin service department for instructions on 
resetting your device’s lamp hours. 

 

Uploading FlexRx Refill Prescriptions 

When specified by your prescribing physician, Daavlin home phototherapy devices are equipped with FlexRx 
exposure limiting software. Each time the lamps are turned on, the number of available exposures is reduced 
by one.  

If your device is equipped with FlexRx, upon “waking up” the screen will display pre-set reminders of how many 
exposures are remaining.  To ensure there are no interruptions in your therapy, we suggest that you contact 
your physician for a “refill” when 5 to 10 exposures remain. Once you have completely run out of exposures on 
your unit, the controller will prompt you to upload a new FlexRx prescription.  

 

UVA 

Narrow Band 

 UVB 

Broad Band  

UVB 

-F72T12BL/HO 

or 

-060OF72HO BL

 

-TL-01/100W-FS72  

or 

-FS72T12/NBUVB/HO/RDC/100W 

-FS72T12HO 

Summary of Contents for 7 Series

Page 1: ...MANU0140 Rev 0 2 2019 1 Operation Manual Integrated Dosimetry with LinkTM Control...

Page 2: ...cking the Device 8 Changing the Key Code 8 Features of the LinkTM Control System and Integrated Dosimetry 8 Operating the LinkTM Home Control System 9 Setting Up a Treatment 9 Care of Your Phototherap...

Page 3: ...unit can be left in place without being in the way of traffic flow It is important that the unit be properly grounded Extension cords are not recommended The location should not be in any area where w...

Page 4: ...foot so that its longer side is toward the front the lamp side of the device Mount it using the screws that you removed from the bottom of the device Repeat the process for the second foot 4 Using two...

Page 5: ...e and if there is any question as to whether or not it is equipped with the lamps that you have been prescribed contact Daavlin immediately The code numbers shown are generally located at the base of...

Page 6: ...g label an identification label indicates the serial number and date of manufacture that is specific to your device A chart detailing all symbols located on your unit can be found on the following pag...

Page 7: ...based sunscreen For your first treatment unless otherwise instructed by your physician do not exceed the following doses Narrowband UVB 200 milliJoules Broadband UVB 20 milliJoules UVA 1 Joule If you...

Page 8: ...trol System and Integrated Dosimetry The Link Controller is an advanced control system which automatically performs many safety checks on your behalf as well as calculating dose increases based on you...

Page 9: ...was in pain or red for 24 48 hours after the treatment More than 48 hours I was in pain or red for more than 48 hours Still red pain I am currently still red or have pain Note Selection of the Still...

Page 10: ...ugh Daavlin When dusting the unit and lamps we recommend using a feather duster or a similar non abrasive cleaning device Note Do not clean reflective surfaces with paper towels as they may scratch th...

Page 11: ...oles in the top of the unit and set it aside 4 Grasp the lamp to be removed with both hands and press down until it clears the top socket then remove the lamp Reverse the process to install lamps and...

Page 12: ...ystems or components as covered in accordance with the terms and conditions of this warranty However this warranty will be declared null and void if non qualified technicians perform any repair or mai...

Page 13: ...ing types of work are examples of maintenance and do not fall within the scope of the warranty Required periodic calibration replacement of lamps cleaning and leveling the equipment Any work performed...

Reviews: