Thimble method example
LEGENDA:
1. Exhaust air from the cabinet.
2. Air from the environment.
3. Bleed Air (100
÷
200 m
3
/h more than the exhaust air
flow rate from the cabinet) to a dedicated exhaust
fan (in the electronic board a voltage free contact is
available –see electrical diagram- to check when
ventilation is ON).
For the connection to the gas and/or vacuum networks, read carefully chapter 2C.
The quantity of heat generated by the cabinet, if the air is not extracted ouside the room, is the following for
the eight models respectively:
The quantity of heat generated by the cabinet, if the air is not extracted outside the room, is the following for
the four models respectively:
-
CytoFAST Top 209:
410 Kcal/h
-
CytoFAST Top 212:
460 Kcal/h
-
CytoFAST Top 215:
520 Kcal/h
-
CytoFAST Top 218:
540 Kcal/h
The installation must be done by technicians authorized by Faster S.r.l. or by the distributor
Microbiological Safety Cabinet
THIMBLE
3
2
1