9 MONITORING SYSTEM
ALARM OR ERROR MESSAGE
DESCRIPTION
Min. LAF Alarm
Airflow speed in the work chamber under minimum threshold value
Min. BARR Alarm
Insufficient air suction from the front aperture
LAF Triac Failure
Too fast and uncontrollable main motor-fan (LAF)
No encoder input LAF
No input from the fan anemometer installed on the LAF main motor-fan
No encoder input EXH
No input from the fan anemometer installed in the exhaust duct
Max LAF Alarm
Airflow speed in the work chamber over maximum threshold value
Front Glass open
The front window is not well closed.
UV Lifetime over
Residual lifetime of the UV lamp expired
Sensor failure
One of the two sensors does not work correctly, press “MUTE” and the
display show the sensor code
BLACK-OUT
Warning of black-out when the main key-switch of the cabinet is in
position “1” (cabinet working).
Press “MUTE” to stop the alarm.
HEPA Filters Check
Possible clogging of absolute (HEPA) filters
Exhaust Duct Check
Possible clogging of the exhaust duct
F1 (2, 3, 4, 5) Lifetime over
Residual lifetime of the Absolute filters (HEPA) expired
AUX Lifetime over
Residual lifetime of the auxiliary filter installed in the exhaust kit (if
installed) expired.
ATTENTION DANGER Reduced
airflow DO NOT WORK
Warning to pay attention when the speed reduction is activated (only for
keyboards with “SPEED REDUCTION” key)
Summary of Contents for FASTER CytoFAST Top
Page 25: ...a D GROUP Company 00055EN Rev 10 12 2011 24 Picture 1 Picture 2 ...
Page 39: ...a D GROUP Company 00055EN Rev 10 12 2011 38 Side view ...
Page 41: ...a D GROUP Company 00055EN Rev 10 12 2011 40 Side diagram ...
Page 42: ...a D GROUP Company 00055EN Rev 10 12 2011 41 13 C SAFE WORKING AREA DIAGRAM ...