7 MAINTENANCE
7.A INSTRUCTIONS FOR DAILY CLEANING OF CytoFAST Top (by users)
Clean the outside of the CytoFAST Top cabinet, made of varnished iron-carbon sheet steel, using a damp
cloth soaked in soapy water or some other commonly available products for varnished metal surfaces.
The recommended procedure to follow is:
1. Switch off the cabinet and disconnect it from the mains;
2. Clean carefully all the painted surfaces using a damp and squeezed cloth with soapy water or equivalent
product.
3. Wipe all the surfaces using a soft cloth absolutely not abrasive.
If the cabinet is used for pathogenic agents is necessary Clean/sterilize the internal work chamber with a
bactericidal agent (e.g. phenolic compositions, aldehydes, quaternary ammonium salts, etc.).
The recommended procedure to follow is:
1. Switch off the cabinet and disconnect it from the mains;
2. Protect your hands wearing gloves if required by the bactericidal agent you are using
3. Spray the bactericidal agent on the side and the back walls and on the work surface in a such uniform
way to avoid leaving not sprinkled zones.
4. Do not spray the filtering upper zone to avoid damaging this part
5. Wipe carefully all the sprinkled zones using clean blotting paper. Be sure to remove any residual of the
bactericidal agent.
If pathogenic material has been spilled during the work session, sterilize the work surface, then remove it
using the small handles mounted on the sides. Sterilize the spilled materials contained in the collection basin
located under the surface and remove all spilled material by means of blotting paper.
For thorough cleaning we suggest the use of solution.
NOTE:
Never use solutions containing free chlorine (for instance, sodium hypochlorite),
which cause corrosion to steel and stainless steel, resulting in irreparable damage to
the cabinet structure.
IMPORTANT: For the procedure of DECONTAMINATION of the work surface, pls refer to the
instructions given by the manufacturer of the cytotoxic drug to be manipulated
Summary of Contents for FASTER CytoFAST Top
Page 25: ...a D GROUP Company 00055EN Rev 10 12 2011 24 Picture 1 Picture 2 ...
Page 39: ...a D GROUP Company 00055EN Rev 10 12 2011 38 Side view ...
Page 41: ...a D GROUP Company 00055EN Rev 10 12 2011 40 Side diagram ...
Page 42: ...a D GROUP Company 00055EN Rev 10 12 2011 41 13 C SAFE WORKING AREA DIAGRAM ...