Cygnus DIVE Operating Manual
M3-DIVE-M-ENG_Iss9.docx
Page 64 of 117
item and use the buttons to switch between Single Echo (SE) or
Echo-Echo (EE) modes.
Probe Type Selection (PROBE TYPE)
The gauge can automatically detect the
following probe types when the probe
type is set to AUTO;
•
S2C Probes
•
T5B Probes
•
T2C Probes
If you have a S3C, S5C or S5A probe
then you will need to manually set the
probe type to match the probe being
used.
To change the probe type select PROBE
TYPE from the Main Menu.
The probe type must be
correctly set to match the probe
being used otherwise thickness
measurements will be impossible or
incorrect.
Probe Type = AUTO
Probe Type = S2C
Setting the Power Save Timer (POWER-SAVE TIMER)
The gauge has a power save feature that
helps to increase battery life when the
gauge is not measuring.
1.
First, after a period of inactivity, the
thickness measurement and A-Scan
graph will be turned off and the
gauge measurement scan rate will
slow down.
2.
After a further period of inactivity the
Power Save Timer
Setup Screen