Cygnus DIVE Operating Manual
M3-DIVE-M-ENG_Iss9.docx
Page 45 of 117
Heavily Corroded Steel Sample.
The best approach when having trouble getting Multiple Echo
measurements is to try slowly moving the probe around the area
looking for points where the echo pulses are stronger and more
defined, keeping an eye on the A-scan so you spot
“better” areas.
If you still are unable to get Multiple Echo measurements then you
have the option of using a twin crystal probe and using the gauge
in Single Echo mode.
12.?mm
UT Probe