Cygnus DIVE Operating Manual
M3-DIVE-M-ENG_Iss9.docx
Page 27 of 117
Automatic Probe Type Detection
The DIVE gauge can automatically determine the probe type when
it is first connected to the gauge setting the appropriate
parameters to suit the Probe Type. This feature will operate when
the probe type is set to AUTO
5
. To set the probe type see Probe
Type Selection on page 64.
Twin Element Probe Wear Indication
The twin element probes have a hard-face that under normal use
should last a long time. However this hard face will gradually wear
down due to the abrasive action of measuring corroded metals.
When the probes hard face has worn down by 2mm (0.08in) or
more the gauge will detect this and display a PROBE WORN
message to the user, the gauge will then not be able to use this
probe for measurements and a new probe must be used.
Probe Worn Message.
Probe Worn Message on Measurement Screen.
5
AUTO will only detect S2C, T5B & T2C Probe Types. Other frequencies have to be manually
set.