Technical Specifications
Cygnus M5-C2P
Technical Specifications
General Attributes
Size
84 mm x 130 mm x 35 mm (W x H x D)
(3.3 in x 5.1 in x 1.4 in)
Weight
Gauge with batteries 300 g (10.5 oz.)
Power Supply
3 x AA / R6 Batteries.
Probe Sockets
2 x Lemo 00
Operating Temperature Range -10°C to +50°C (14°F to 122°F)
Storage Temperature Range
-10°C to +35°C (14°F to 95°F)
Battery Operation Time
Approximately 10 hrs. continuous measurement with fully
charged Duracell Alkaline LR6 batteries.
Battery Voltage Range
Min 3.0 V dc, Max 4.5 V dc
Battery Type
AA Size. LR6 Alkaline / HR6 NiMH.
Low Battery Indication
Battery level indication on display with low battery warning
message.
PRF
N/A
Monitor Outputs
N/A
Through Coating
Measurements
Multiple-Echo mode with Single element 0° probe;
•
Through coating measurement for coatings up to 3 mm
thick as standard depending on coating velocity.
•
Deep Coat mode provides ability to measure through
thicker coatings depending on coating material.
Echo-Echo mode with Twin element probe;
•
Through coating measurement for coatings up to 1 mm
thick as standard depending on coating velocity.
Materials
Sound velocity from 1000 m/s to 9000 m/s
[0.0390 in/us to 0.3543 in/us]