M5-CYG2P-M-01_Iss6.doc
Cygnus 2+ Operating Manual
41
3.
Press the Menu key to change
the probe type
4.
Use the Up and Down Select
keys to select from Auto or a
specific probe type
5.
When done press the Menu key
to save the new value.
Or press the Power / Cancel to
exit without saving.
Remember settings and calibration is saved with each probe
type. So if you change the probe type you may find a setting
has changed – this is normal.
The reason settings are saves with each probe type is so you
can swap probes during a survey without having to re-
calibrate each time.
Also you may want Multiple Echo measurements in 0.1mm
resolution but Single Echo measurements in 0.01mm
resolution.
Automatic Probe Detection
The Automatic Probe Detection works by listening to how the
probe behaves when first plugged in. For that reason the probe
MUST BE in the air and not coupled to a surface when plugging in
or turning on the gauge.