40
GenesisPlus Operator Manual
D1151, R
EV
C, 10/16
Removable or Lockable Wall Socket and Plug Configurations
If your system is installed with a removable plug or wall socket and lockable plug configura-
tion prior to installation, the customer’s engineer or electrical contractor will be responsible for
ensuring that the proper electrical requirements are available at the site.
Systems Designed for Use in European Communities Under MDD
To comply with the European Communities Medical Devices Directive 93/94/EEC, the 16A,
250V wall socket and lockable plug combination must comply with EN 60309 (=IEC 309).
The wall socket and plug combination must meet the following specifications:
Plug:
MK LN 9024, 16A 220V, IP67, IEC 309
Wall Socket:
MK LN 9324, 16A 220V, IP67, IEC 309