13
GenesisPlus Operator Manual
D1151, R
EV
C, 10/16
In READY mode:
• the footswitch is enabled
• the safety shutter is open
• the treatment, aiming and working distance beams are available
• the temperature sensor and LED temperature indicators are functional
• the
(ready) button and the borders around the treatment parameter displays are
highlighted in green
Press to go to
STANDBY mode
To return to STANDBY mode, press the
(standby) button.
When the system is in READY mode and the footswitch is depressed, treatment energy is dis-
charged from the handpiece, the system emits an audible tone and a yellow indicator next to
the
(ready) button illuminates to indicate laser emission.
If the system is not used for more than three
minutes, the system automatically goes into
STANDBY mode.
WARNING
Except during actual treatment, the system must always be in
STANDBY mode. Maintaining the system in STANDBY mode
prevents accidental treatment beam exposure if the footswitch is
inadvertently depressed.
WARNING
Verify that all persons in the treatment room are wearing the appro-
priate treatment beam safety eyewear before placing the system in
READY mode.
Summary of Contents for GenesisPlus
Page 1: ...i GenesisPlus Operator Manual D1151 REV C 10 16 Operator Manual ...
Page 6: ...vi GenesisPlus Operator Manual D1151 REV C 10 16 THIS PAGE IS BLANK ...
Page 70: ...64 GenesisPlus Operator Manual D1151 REV C 10 16 ...
Page 72: ...A 2 APPENDIX A GenesisPlus Operator Manual D1151 REV C 10 16 THIS PAGE IS LEFT BLANK ...
Page 78: ...C 2 APPENDIX C GenesisPlus Operator Manual D1151 REV C 10 16 THIS PAGE IS BLANK ...
Page 80: ...D 2 APPENDIX D GenesisPlus Operator Manual D1151 REV C 10 16 THIS PAGE IS LEFT BLANK ...