EZT-570i Technical Manual
38
SYMPTOM PROBABLE
CAUSES
CORRECTIVE
ACTIONS
alarm message.
and control module not properly
connected.
Electrical noise in system or remote PC
communicating too fast with EZT.
Correct/repair.
Check for proper shielding of
communication wiring.
Turn off electrically noisy devices.
Disconnect remote PC or slow down
communication rate from remote PC.
(minimum 500ms scan rate)
4) Screen Locked Up
Screen does not respond to touch.
Touch screen out of calibration.
VNC server not properly disconnected
after previous access via remote PC.
HMI operating temperature exceeded.
CE operating system locked up.
Touch screen failure.
Calibrate touch screen.
(see Section 6.1.3)
Restart VNC client at PC and verify
operation. Shutdown VNC client to
regain touch screen operation
Ambient temperature around HMI
must be between 0°C (32°F) and
45°C (113°F). Turn off EZT. Bring
ambient temperature to within
allowable range.
Cycle power.
Replace HMI.
Summary of Contents for EZT-570i
Page 1: ...EZT570i Technical Manual Rev B EZT 570i Technical Manual...
Page 14: ...EZT 570i Technical Manual 14 4 1 6 Conditioning System Logic Flow...
Page 15: ...EZT 570i Technical Manual 15...
Page 19: ...EZT 570i Technical Manual 19 4 2 2 Humidity System Logic Flow...
Page 21: ...EZT 570i Technical Manual 21 4 3 2 Auxiliary Cooling System Logic Flow...
Page 23: ...EZT 570i Technical Manual 23 4 4 2 Dry Air Purge System Logic Flow...
Page 25: ...EZT 570i Technical Manual 25 4 5 2 Altitude System Logic Flow...
Page 29: ...EZT 570i Technical Manual 29 4 7 2 Transfer System Logic Flow...
Page 30: ...EZT 570i Technical Manual 30...
Page 58: ...EZT 570i Technical Manual 58 7 2 Standard Output Configuration...
Page 92: ......