4.1.5 Conditioning System Failures and Corrective Actions
SYMPTOM PROBABLE
CAUSES
CORRECTIVE
ACTIONS
Chamber air circulator(s) not turning
on.
Chamber event not turned on.
Conditioning system disabled (altitude
chambers).
Blown fuse.
Chamber in defrost (if equipped).
Basket not in correct position
(DTS/VTS).
Chamber door open (if door switch
enabled in configurator).
Turn on chamber event.
Altitude above controllable limit for
temperature. Decrease altitude or
turn off altitude system.
Replace Fuse.
Check defrost status. Allow defrost to
complete or terminate defrost.
Check basket position and adjust if
necessary. Check basket position
sensors, adjust/replace.
Close chamber door.
Chamber not heating.
Chamber event not on.
Conditioning system disabled (altitude
chambers).
Blown heater fuse.
Sheath heater limit tripped.
9300 controller output off.
Chamber in defrost (if equipped).
Turn on chamber event.
Altitude above controllable limit for
temperature. Decrease altitude or
turn off altitude system.
Replace fuse.
Check heater limit. Maximum
temperature reached.
Verify proper 9300 controller
configuration. Check set point.
Replace 9300 controller.
Check defrost status. Allow defrost to
complete or terminate defrost.
System 1 compressor not turning on.
Chamber event not on.
Conditioning system disabled (altitude
chambers).
Blown fuse.
Compressor internal thermal overload
tripped.
Chamber in defrost (if equipped).
Refrigeration system not enabled or
compressor percent on set point not
exceeded for delay time.
Turn on chamber event.
Altitude above controllable limit for
temperature. Decrease altitude or
turn off altitude system.
Replace fuse.
Allow compressor to cool. Check
refrigeration system/injection valve
operation.
Check defrost status. Allow defrost to
complete. Compressor will cycle as
needed.
Check configurator settings. Refrig
system type should match installed
system type.
System 2 compressor not turning on.
Blown fuse.
Replace fuse.
Summary of Contents for EZT-570i
Page 1: ...EZT570i Technical Manual Rev B EZT 570i Technical Manual...
Page 14: ...EZT 570i Technical Manual 14 4 1 6 Conditioning System Logic Flow...
Page 15: ...EZT 570i Technical Manual 15...
Page 19: ...EZT 570i Technical Manual 19 4 2 2 Humidity System Logic Flow...
Page 21: ...EZT 570i Technical Manual 21 4 3 2 Auxiliary Cooling System Logic Flow...
Page 23: ...EZT 570i Technical Manual 23 4 4 2 Dry Air Purge System Logic Flow...
Page 25: ...EZT 570i Technical Manual 25 4 5 2 Altitude System Logic Flow...
Page 29: ...EZT 570i Technical Manual 29 4 7 2 Transfer System Logic Flow...
Page 30: ...EZT 570i Technical Manual 30...
Page 58: ...EZT 570i Technical Manual 58 7 2 Standard Output Configuration...
Page 92: ......