data logger under cycle B etc., so devices with more than one winding can be tested. The
following screen is displayed after pressing the STOP button twice, or one the MAX time
(END TIME) has elapsed.
After double pressing of the STOP key or after max. time (END TIME) you get the following
display:
With the
↑
arrow button you can view the values.
Use the
↑↓
buttons to view the measured values.
The REL TIME is the time elapsed after pressing LOAD REM.
If you have selected PRINTER as the data output device, you can now print out the table in full.
If you have selected PC as the data output device, you can now transfer the values to the PC
via the interface.
Since the first resistance value cannot be measured until after a short delay after switching off
the load current; the actual resistance at the time when the load was removed can only be
found by extrapolating the cooling curve. The add-on PC software package DO6-P001 can be
purchased to help perform this calculation.