15
At the end of the depressurization, the drying stage will begin automatically, the display will show the message
drying
and the temperature of the chamber.
The autoclave Quadra 54 dries with the door closed
. At the end of
the drying process, the autoclave will beep 3 times and the display will show the message
cooling
. At this moment,
the operator must open autoclave door and leave it ajar to accelerate the cooling process of the sterilized articles.
Within 5 minutes the autoclave will beep continuously and the display will show the message
cycle concluded
, only
then, remove the sterilized articles.
After cooling, if the flash drive memory unit is not connected, the equipment will beep and request it, by showing
the message
on the display.
insert DISk or cancel
Note:
In case the operator decide not to use the flash drive by
pressing
Cancel
, the display will show the message:
transfer canceled
, the data will not be transferred and there will
be no way to recover the information from that cycle later on.
f
g
While the data is being transferred, the display will show
transferring...
. In case the flash drive is full or there’s a
transfer error, the autoclave will alert the user about it. If the user needs to format the flash drive, it must be formatted as
a FAT file system, FAT32 system is not supported by the electronic circuit board of the equipment. When the data
transfer is finished, the display will show
CYCLE concluded
and it will beep until the
CANCEL
key is pressed. The autoclave
will then return to
Standby
mode and the display will show
SELECT PROGRAM
.
To transfer the data from the flash drive to your computer, simply insert it into the USB connection on your computer
and copy the reports to a local folder and print them.
CRISTÓFOLI
A U T O C L
A V E S
I N S T R U C T I O N M A N U A L
e
Select the cycle desired by pressing the
MODE
key. Each cycle has a specific operation time and temperature. Each
time the
MODE
key is pressed, the display will show the preset programs in the following order:
Pressing once:
wrapped Instruments
10 min.
Pressing twice:
Plastics AND COTTON
30 min.
Pressing three times:
SURGICAL Kit and fabrics
30 min.
Pressing four times:
Liquids
30 min.
Pressing five times:
additional info
(see “
Cycles and Messages
”, page 13)
d
After choosing one of the preset cycles through the
MODE
key, press the
START
key;
In case this is not the first cycle of the day and the temperature of the autoclave is above 70 ºC, the display will
show
cooling
and the respective temperature. As soon as the ideal temperature for the new cycle is reached, the
autoclave will resume operation automatically by changing to
and then
heating UP
;
WATER filling
, the display will show the temperature and pressure rising gradually.
When the solenoid valve shuts down
When
the autoclave reaches the ideal sterilization temperature and pressure for the selected program, it will beep 3 times
and the display will show
steriliz.
(sterilizing) remaining like that for the preset time according to the program
chosen.
At the end of the sterilization, the solenoid valve will open, it will produce a low metallic noise, the autoclave will
beep 4 times and the display will show
depress.
(depressurizing).
During the heating and drying stages, the solenoid valve makes a humming noise, similar to the one produced
by electric motors, it may also open/close automatically during the drying stage.