Check if the power cable is plugged in, switch on the circuit breaker in the back of the autoclave, the display will 

show the following mesages:                        and                     , next, the display will go out. In order to turn the 

display on again, press the 

START

 key for 2 seconds, the display will show the initial information (brand, model and 

software version), the autoclave will beep twice and the display will show 

select program

.

bootloader V000

aguarde . . .

bootloader V000

executando . . .

HOW TO USE THE AUTOCLAVE QUADRA 54

a

SAFETY NOTES

b

Open the door of the autoclave and load it with the materials to be sterilized, be careful not to lean them against 
the chamber walls or internal steam exits, that will cause interference in the cycle and damages to the articles. 

Do 

not overload the autoclave

.

 

Warning!

  -  During  the  autoclave  operation,  it  is  perfectly  normal  to  hear  some  noises,  some  low,  some  loud.  The 

noises are generated by the valves opening and closing, deaeration and depressurization, they are part of the proper 
functioning  of  the  equipment.  The  door  gasket  and  the  VSPF  valve,  are  safety  mechanisms  that  when  activated 
discharge  pressure  automatically  producing  a  loud  noise.  This  autoclave  must  be  installed  in  a  proper  and  exclusive 
sterilization  room.  Cristófoli  is  not  responsible  for  accidents  that  might  occur  due  to  the  starts  caused  by  the  noises 
produced by the autoclave.

 

The symbol 14     appears in some places on the autoclave. This means those items require special attention and 

that the user must observe their references in the Instruction Manual provided with the equipment regarding potential 
hazards offered by them and any actions to be taken should an adverse situation occur.

 

Cristófoli  Equipamentos  de  Biossegurança  Ltda,  does  not  take  any  responsibility  for  failures  and/or  accidents 

caused by the non-observance of this warning.

ATTENTION! 

Before starting a new sterilization cycle, make sure the reservoir has enough water to supply the autoclave need 

(at least 1,5 liters). 

Use only distilled water for the sterilization 

(in accordance with the EN 13060)

.

 Non-compliance with this 

recommendation may cause obstructions of the autoclave hydraulic system (internal tubing and/or valves), stains on the 
instruments and loss of warranty.

Close the door of the autoclave. To close it correctly, press it against the chamber, slide the handle downwards until 
it reaches the end of the course (Fig. 6, page 11). To open the autoclave, slide the handle upwards and open the 
door.

ATTENTION!

 In case the door of the autoclave is not properly closed, the program will not start. The autoclave will beep 

and the display will show the message: 

close the door!

Summary of Contents for Autoclave Quadra 54

Page 1: ...equipment to assist the health field having as policy Develop innovative solutions for the health field by using agile robust and objective processes to better serve its clients Fulfill the requiremen...

Page 2: ......

Page 3: ...ance in the coordination of Crist foli s Biosafety Project complementary products research training of our employees representatives and technicians and as a lecturer of Biosafety Courses for professi...

Page 4: ...oclave Time x Pressure Graphs Technical Data Preventive Maintenance Preventive Maintenance Table Troubleshooting How to Proceed When Service is Needed Guidance for Final Disposal of the Equipment Warr...

Page 5: ...erminal Fusible Autoclavable Date of manufacture Boas Pr ticas de Fabrica o 4 P B F LOT SN Fragile Handle with care Dangerous Electrical Tension Serial Number Batch code Keep away from sunlight Recycl...

Page 6: ...for sterilization we recommend extra attention to this manual Autoclave sterilization although quicker and more efficient requires totally different procedures and handling while still keeping a simp...

Page 7: ...JAN 01 2009 The piece of information to be set will be flashing 3 Press MODE or CANCEL to set the data 4 Press START quickly to alternate the target information day month and year and keep it pressed...

Page 8: ...tem b page 14 When sterilizing simultaneously instruments that have different autoclaving specifications as instruments that support 2 different levels of pressure 1 2 or 2 1 kgf cm or temperature 121...

Page 9: ...red It is mandatory the use of an exclusive circuit breaker for the outlet where you intend to install your autoclave If after following all the specifications the electrical current is still oscillat...

Page 10: ...ification Label Power Cable Circuit Breaker Power Supply Buzzer V cut Hose for the external steam exit Keep it attached to the lid above the water level Use a lid with holes so the steam flows out Res...

Page 11: ...rs each It s located on the keyboard Fig 5 HANDLE Located in the right side of the autoclave panel Figs 8 and 9 page 11 It s used to open close and lock the autoclave door panel lid set To check the c...

Page 12: ...nents of the equipment responsible for the deareration depressurization and water supplying The valves open the beginning of the heating stage eliminating the cold air from the chamber then it shuts d...

Page 13: ...the equipment to the power outlet Fig 1 page 9 CLAMP Used to fasten the discharge hose to the external steam exit Fig 12 page 11 BUZZER Internal device used is to emit the beeps produced by the autoc...

Page 14: ...how many times the user has been notified about performing the biological test 5 Shows how many days ago and how many times the user has been notified about performing the preventive maintenance 6 Sho...

Page 15: ...alled in a proper and exclusive sterilization room Crist foli is not responsible for accidents that might occur due to the starts caused by the noises produced by the autoclave The symbol 14 appears i...

Page 16: ...from the flash drive to your computer simply insert it into the USB connection on your computer and copy the reports to a local folder and print them CRIST FOLI A U T O C L A V E S I N S T R U C T I...

Page 17: ...ry to speed up the cooling process of the instruments articles sterilized Upper view Note Heating time values are expressed considering the technical data chart Table 5 page 19 regarding the autoclave...

Page 18: ...when the ideal pressure temperature is not reached due to steam pressure leak or overloaded chamber In this case the cancellation will occur within 50 minutes at the most the sterilization cycle is i...

Page 19: ...N The removal of the identification label and or any other labels stickers affixed to the product will cause automatic loss of warranty Note The label presented here is just a sample model for your re...

Page 20: ...ce required for the movement of the door Chamber internal dimension W x D Autoclave external dimension W x H x D Dimensions of the trays W x H x D Clearance between the trays Voltage Frequency Power P...

Page 21: ...y Perform the preventive maintenance on the VSPF valve annually it consists of replacing the internal safety valve that comprehendsthe seal and pin Check the door locking system of the autoclave annua...

Page 22: ...L MAINTENANCE DAILY WEEKLY ANNUAL EVERY 10 YEARS External cleaning Preventive cleaning stainless steel chamber Door gasket cleaning Door gasket replacement Preventive Maintenance of the VSPF valve and...

Page 23: ...OT SWITCH ON THE AUTOCLAVE PRESSURE GETS TOO HIGH ACTIVATING THE SAFETY DEVICES Partial obstruction of the solenoid valve Defective electronic circuit Obstruction of the hose connected to the external...

Page 24: ...ance Page 20 Check the water level in the reservoir if necessary fill it up to the correct level Do not put more instruments than specified in each bag package Use up to 75 of the chamber s capacity t...

Page 25: ...ive Maintenance Item 8 page 20 Crist foli Equipamentos de Biosseguran a Ltda makes clear its concern already demonstrated by the implementation of the Environmental Management System according to stan...

Page 26: ...ssive exposure to sun light and salinity Sinister theft or robbery Improper use installation of the equipment and or plugging it into an incorrect voltage outlet Problems due to failure of the power s...

Reviews: