2.2 Operation
The Ultrasonic Cleaning System uses ultrasonic energy to
thoroughly wash by removing tissue, blood and other
contaminants or dry laparoscopy and general surgical instruments
prior to final disinfection and sterilization as required. Please refer
to the latest Intuitive Surgical Reprocessing Instructions for
detailed descriptions of instruments and accessories, designed for
use with Intuitive Surgical’ s
da Vinci Surgical
Systems and which
are currently validated for use with this equipment. When
processing Intuitive Surgical instruments, it is important that the
healthcare facility follow all directions in the latest Intuitive Surgical
Reprocessing Instructions as well as any instrument specific
instructions and supplements that may come with individual
instruments (presoaking steps should precede; refer to Intuitive
Surgical IFUs).
Operation other than as specified and described in this section is not
recommended, may not be effective in cleaning instruments and
may not be safe. Please contact Crest for appropriate guidance, in-
service and training. For custom Cycle set-up, contact Crest or a
properly trained Service Technician.
If initial start-up of a Ultrasonic Cleaning System, proceed to
S
ECTION
5.4.1, I
NITIAL
S
TART
-
UP
. If not initial start-up of a Ultrasonic
Cleaning System, proceed to next section.
2.2.1 Pro Cleaner Cycle, Pre-
Wash through Final Spray
Phases
After Initial Start-Up, use facility safe instrument handling practices to
load Instrument Trays and process load as shown in
F
IGURE
2-1
.
NOTE: For Cleaning System Cycle operation, note the following:
1) If Operator is running Cycles 4/5 and then chooses Cycle 1/2/3,
Cleaning System automatically runs Cycle 6 flush and drain
(unless REMAINING WASH STEP is at 0 and system has power
drained). Lid opens at Cycle 6 completion and Operator must
reselect Cycle 1/2/3.
2) If Operator is running Cycles 1/2/3 and then chooses Cycle 4/5,
Cleaning System drains chamber and Cycle 4/5 begins.
3) If Operator is running Cycles 4/5 and Cleaning System reaches
Cycle count, if Operator selects another Cycle 4/5, Cleaning
System drains chamber between Cycle selections.
2-2
Operator Manual
Operating Instructions
WARNING – PINCHING
HAZARD: Pinch point is
created during lid articulation.
Keep hands clear when
opening or closing the lid
.
CAUTION – POSSIBLE
EQUIPMENT DAMAGE: Avoid
running equipment into Pro
Ultrasonic Cleaning System
Control Panel or Touch
Screen to prevent damage to
ontrols.