Service Manual

16

VNUS Radio Frequency Generator

Service and Maintenance

Cleaning and Disinfection

 Caution: Always unplug the AC Power cord prior to cleaning the RF Generator.  Avoid spilling liquids into the 

RF Generator during cleaning.  Avoid getting cleaning materials inside RF Device receptacle.

The RF Generator requires no scheduled maintenance other than cleaning external surfaces . It is 

recommended that the operator clean the unit prior to each use or as needed . 
The recommended agents for cleaning RF Generator surfaces (not including the LCD screen) to prevent 

surface degrading or discoloration include the following:

  •  Five percent solution of household bleach (approximately 2,500 ppm Sodium Hypochlorite)
  •  Sporicidin®
  •  IPA 70/30

 
The recommended agent for cleaning the transparent, protective cover of the LCD screen is IPA 70/30 .
RF Generator cannot be sterilized and must not enter a sterile surgical field .

Repairs and Returns

The RF Generator contains no operator-serviceable parts . The RF Generator must be either returned to 

the factory or repaired by qualified service personnel . 
For RF Generator repairs or returns, call the VNUS Customer Service Department . See the back of this 

manual for contact information . A Customer Service Representative will assist the operator with 

obtaining service or returning the RF Generator, as appropriate .
The RF Generator must be unplugged before servicing .

WARNING: Do not remove the cover of the RF Generator, as there is a potential for electrical 

shock . 

Warranty

For information about the RF Generator warranty and/or to obtain information about the availability of 

the VNUS Extended Warranty, contact VNUS Customer Service . See the back of this manual for contact 

information .
Opening the RF Generator by anyone other than an authorized VNUS technician voids the warranty .

Fuse Replacement

The RF Generator uses two 4A/250V, slo-blo, 5x20 mm fuses . In the event of a fuse failure, replace the 

fuses using the following procedure, as shown in Figure :

 Caution: Fuses must be replaced with 4A/250V, slo-blo, 5x20 mm fuses. Using other fuses might damage 

the unit.

 1 .  Disconnect the RF Generator’s AC Power cord from the AC Power outlet .
 2 .  Remove the slide-in fuse holder by inserting a small blade screwdriver into the slot located on the 

bottom side of the holder while activating the latch, and then pull out the holder by hand until the 

holder slides out . 

 3 .  Remove the holder completely by hand and replace the fuse(s) .
 4 .  Reinsert the fuse holder by pushing it with the thumb until the holder’s front surface is flush with the 

power entry module’s surface and the latch engages .

 

 

 

Fuse

holder

5 x20 mm

fuses

AC Power

switch

AC Power

inlet

Latch

 

Figure 6: Fuse Replacement

Summary of Contents for VNUS RFG2

Page 1: ...VNUS Radio Frequency Generator RFGPLUS Model RFG2 Service Manual...

Page 2: ...iance 8 Biomedical Department Power OutputTest 12 Power DeliveryVerification 13 Alarm Priority 13 Error Codes 14 Service and Maintenance 16 Cleaning and Disinfection 16 Repairs and Returns 16 Warranty...

Page 3: ...ator The RF Switch circuitry contained on the UI Board provides the switch and colored backlighting for the RF on off switch located on the right of the FPI as viewed from the front of the Generator R...

Page 4: ...nly be removed with a tool Service Port is UL1950 50VDC Normal MAINSV in Single Fault Condition Metal Enclosure is Protective Earth Display has no touch screen and sits behind protective plastic cover...

Page 5: ...TL100 119 02 1 Set the line voltage to 264V and set the frequency to 50Hz DO NOT EXCEED 280V 2 Plug the line cord of theVNUS Generator into the leakage tester for all tests 3 For source leakage and i...

Page 6: ...tion Nearby electronic equipment may be affected RF emissions CISPR 11 Class A The RFG2 is suitable for use in all establishments other than domestic and those directly connected to the public low vol...

Page 7: ...e considered If the measured field strength in the location in which the RF Generator is used exceeds the applicable RF compliance level above the RF Generator should be observed to verify normal oper...

Page 8: ...nce thermometer and the ESU analyzer Turn off main power to the generator Remove the modified Instrument Cable from the RF Generator and AC power cord from the generator In assessing the functionality...

Page 9: ...es Turn off the unit wait a few seconds and then turn on the unit If the error persists try a different MMC Error 26 The inserted Multimedia Card MMC cannot upgrade this RF Generator because its seria...

Page 10: ...tative will assist the operator with obtaining service or returning the RF Generator as appropriate The RF Generator must be unplugged before servicing WARNING Do not remove the cover of the RF Genera...

Page 11: ...e SJT16 3 IEC color code maximum length 10 ft 3 m Plug minimum 10 A 125VAC Unit receptacle IEC female minimum 10 A 125VAC 220 240VAC Cable H05VVF3G1 0VDE maximum length 10 ft 3 m Plug minimum 6 A 125V...

Page 12: ...0 10 30 40 50 60 70 80 90 100 0 5 10 15 20 25 30 35 40 Peak Voltage V Power W Figure 8 Peak Output Voltage vs Power Output 0 0 5 5 10 10 15 15 20 20 25 30 35 40 25 30 35 40 Power Output W Max Power Se...

Page 13: ...88 797 VEIN 8346 Fax 1 408 365 8470 Covidien Ireland Limited IDA Business Technology Park Tullamore 0123 Customer Service Tel 1 800 962 9888 MEDICAL EQUIPMENT MODEL RFG2 WITH RESPECT TO ELECTRICAL SHO...

Reviews: