Service Manual

6

VNUS Radio Frequency Generator

During the Power On Self Test (POST), the RF Generator performs a number of internal self-tests of its 

software:

  •  Verifies integrity of system random access memory (RAM).
  •  Verifies contents of program read-only memory (ROM).
  •  Activates all visual indicators, including the LCD screen and the backlit 

Soft Key Menu

 buttons 

around the screen .

  •  Sounds Power On tone. 
  •  Verifies operation of radiofrequency (RF) generation and measurement circuits.
  •  Displays a notice if POST must restore the factory default settings due to a corruption of persistent 

settings, such as screen brightness or tone volume settings .

Prior to using the RF Generator, observe that the screen, lights, and audible alarms cycle .

Software Version

The RF Generator’s software release version will be displayed when the Power On Self Test is completed .  

The RFG2 enters Identification Mode as shown in Figure 5 below . 

RFG2 - VERSION 4.1.0

Figure 5: Identification Mode Showing Software Version

Electrical Safety Tests

The VNUS RF Generator has been thoroughly tested for isolation, leakage, and grounding integrity prior 

to shipment . Should it become necessary to perform these tests, the use of the appropriate test 

equipment, and knowledge of the appropriate sections of BS EN 60601-1:1990 +A1 +A2 which pertain 

to these tests is mandatory . VNUS Medical Technologies assumes no responsibility for damage to the 

instrument due to improper application of the tests below . If in doubt, consult the Safety Analyzer Users 

Manual or VNUS Medical Technologies prior to conducting the following tests .

Electrical Safety Tests 220-240V

A special test cable from VNUS is available for these tests. Please contact VNUS Customer Service for 
information on obtaining the Biomed Test Cable (TL100-119-02).

 

1 .  Set the line voltage to 264V and set the frequency to 50Hz . DO NOT EXCEED 280V!

 2 .  Plug the line cord of the VNUS Generator into the leakage tester for all tests .

 3 .  For source leakage and isolation tests, use the Biomed Test Cable to access patient connections . Short 

electrodes and thermocouple wires together for testing .

 4 .  For the 25A test, connect the ROD-L leads to the chassis ground lug, and to the ground pin of the 

power cord .

WARNING! HIGH VOLTAGE!  DO NOT TOUCH THE EQUIPMENT DURING THE ISOLATION TESTS!

Table 2: Electrical Safety Test Limits 240V

Parameter

Conditions

Passing

Input Voltage

Rated line + 10%

 264V

Line Cord Resistance

N/A

< 100mΩ

25A chassis-cord gnd

10 seconds

< 100mΩ

Earth Leakage

Power Norm, GND OPEN

< 500 μA

Earth Leakage

Power Rev, GND OPEN

< 500 μA

Source Leakage, Applied Part

Power Normal, GND OPEN

< 10 μA

Source Leakage, Applied Part

Power Reversed, GND OPEN

< 10 μA

Iso Test, Applied Part

Power Normal

< 10 μA

Iso Test, Applied Part

Power Reversed

< 10 μA

Electrical Safety Tests 110-120V 

 1 .  Use the variable autotransformer and the DVM to set the line voltage to 132V .  DO NOT EXCEED 140V!
 2 .  Plug the line cord of the VNUS Generator into the leakage tester for all tests .
 3 .  For source leakage and isolation tests, use the Biomed Test Cable to access patient connections .  Short 

electrodes and thermocouple wires together for testing .

 4 .  For the 25A test, connect the ROD-L leads to the chassis ground lug, and to the ground pin of the 

power cord .

WARNING! HIGH VOLTAGE!  DO NOT TOUCH THE EQUIPMENT DURING THE ISOLATION TESTS!

Table 3: Electrical Safety Test Limits 120V

Parameter

Conditions

Passing

Input Voltage

Rated line + 10%

132V

Line Cord Resistance

N/A

< 100mΩ

25A chassis-cord gnd

10 seconds

< 100mΩ

Earth Leakage

Power Norm, GND OPEN

< 300 μA

Earth Leakage

Power Rev, GND OPEN

< 300 μA

Source Leakage, Applied Part

Power Normal, GND OPEN

< 10 μA

Source Leakage, Applied Part

Power Reverse, GND OPEN

< 10 μA

Iso Test , Applied Part

Power Normal

< 10 μA

Iso Test, Applied Part

Power Reversed

< 10 μA

Summary of Contents for VNUS RFG2

Page 1: ...VNUS Radio Frequency Generator RFGPLUS Model RFG2 Service Manual...

Page 2: ...iance 8 Biomedical Department Power OutputTest 12 Power DeliveryVerification 13 Alarm Priority 13 Error Codes 14 Service and Maintenance 16 Cleaning and Disinfection 16 Repairs and Returns 16 Warranty...

Page 3: ...ator The RF Switch circuitry contained on the UI Board provides the switch and colored backlighting for the RF on off switch located on the right of the FPI as viewed from the front of the Generator R...

Page 4: ...nly be removed with a tool Service Port is UL1950 50VDC Normal MAINSV in Single Fault Condition Metal Enclosure is Protective Earth Display has no touch screen and sits behind protective plastic cover...

Page 5: ...TL100 119 02 1 Set the line voltage to 264V and set the frequency to 50Hz DO NOT EXCEED 280V 2 Plug the line cord of theVNUS Generator into the leakage tester for all tests 3 For source leakage and i...

Page 6: ...tion Nearby electronic equipment may be affected RF emissions CISPR 11 Class A The RFG2 is suitable for use in all establishments other than domestic and those directly connected to the public low vol...

Page 7: ...e considered If the measured field strength in the location in which the RF Generator is used exceeds the applicable RF compliance level above the RF Generator should be observed to verify normal oper...

Page 8: ...nce thermometer and the ESU analyzer Turn off main power to the generator Remove the modified Instrument Cable from the RF Generator and AC power cord from the generator In assessing the functionality...

Page 9: ...es Turn off the unit wait a few seconds and then turn on the unit If the error persists try a different MMC Error 26 The inserted Multimedia Card MMC cannot upgrade this RF Generator because its seria...

Page 10: ...tative will assist the operator with obtaining service or returning the RF Generator as appropriate The RF Generator must be unplugged before servicing WARNING Do not remove the cover of the RF Genera...

Page 11: ...e SJT16 3 IEC color code maximum length 10 ft 3 m Plug minimum 10 A 125VAC Unit receptacle IEC female minimum 10 A 125VAC 220 240VAC Cable H05VVF3G1 0VDE maximum length 10 ft 3 m Plug minimum 6 A 125V...

Page 12: ...0 10 30 40 50 60 70 80 90 100 0 5 10 15 20 25 30 35 40 Peak Voltage V Power W Figure 8 Peak Output Voltage vs Power Output 0 0 5 5 10 10 15 15 20 20 25 30 35 40 25 30 35 40 Power Output W Max Power Se...

Page 13: ...88 797 VEIN 8346 Fax 1 408 365 8470 Covidien Ireland Limited IDA Business Technology Park Tullamore 0123 Customer Service Tel 1 800 962 9888 MEDICAL EQUIPMENT MODEL RFG2 WITH RESPECT TO ELECTRICAL SHO...

Reviews: