background image

13

Service Manual

12

VNUS Radio Frequency Generator

Biomedical Department Power Output Test

VNUS has created a Biomedical Department Inspection Procedure for the testing of the VNUS RF 

Generator . This procedure outlines setting the VNUS Radiofrequency Generator up in concert with an ESU 

analyzer to verify temperature measurement, power output, and measurement of impedance .
A special test cable from VNUS is provided for this purpose . Please contact VNUS Customer Service for 

information on obtaining the Biomed Test Cable. (TL100-119-02 Cable, Biomed Test).

Set-up Procedure

  •  Check the RF Generator case exterior .  The case exterior shall have no visible damage (e .g . dents, 

scratches, scaled off paint) .

  •  Panel displays shall have no visible foreign material or damage .
  •  Inspect all external connectors/receptacles for bent pins, foreign objects, damaged case and/or 

improper fit .

  •  The AC power cord shall have no obvious physical defects or damage .
  •  Connect the RF Generator to the appropriate AC power supply and turn on main power,  

wait until the generator completes its self- test .

  •  Attach test cable to the generator . 
  •  The front panel parameters should be: 

Temperature:  

~23°C (ambient temperature)

Impedance:  

1999 Ω

Power:   

0 W

  •  Clip ESU tester to red and black leads .  Select 100 Ω impedance on ESU analyzer .
  •  Place red and yellow thermocouple wire in 37°C bath .  Place reference thermometer in 37°C bath .

Test Method

  •  Verify that the bath temperature and ESU analyzer load impedance are indicated on the Temperature 

and Impedance displays .

  •  Depress the “RF Power” switch to initiate RF power sequence .
  •  Verify that the bath temperature, ESU analyzer load impedance, and power indicated on the 

Temperature, Impedance, and Power displays on the RF Generator are in agreement with the 

reference thermometer and the ESU analyzer .   

  •  Turn off main power to the generator .  Remove the modified Instrument Cable from the RF Generator, 

and AC power cord from the generator .

  •  In assessing the functionality of the RF Generator with any external measurement system such as an 

ESU Analyzer, the measurement tolerances of the ESU Analyzer for a given test must be added to the 

tolerances listed in the Tolerances section of this manual .

Power Delivery Verification 

This test is intended to verify the ability of the VNUS RF Generator to deliver RF energy .

  •  Attach a Closure

FAST

 catheter (model CF7-7-60 or CF7-7-100) to the generator .  

  •  Fill a glass container with 0.9% saline solution heated to 37 ±3 °C (this simulates the temperature 

and impedance of blood in a blood vessel) .

  •  Place the distal end of the catheter in saline such that the entire RF coil is covered by the saline.
   •  Press the RF Power button on the generator to enter ready mode .    
  •  Press the Start RF button on the catheter to start RF treatment .  Insure that the catheter coil is 

covered by the saline while RF power is being delivered .  

  •  Verify that the generator delivers power and that the setpoint temperature is maintained during 

delivery .

Alarm Priority

The RF Generator will generate a tonal alarm for an alarm condition based on the power, temperature, 

and impedance feedback .  
The RF Generator will enforce an alarm priority whereby an alarm message or tone takes priority over any 

other type of message or tone (i .e . advisory and informational) .
During multiple message or tone conditions, the message or tone with the highest priority will take 

precedence as shown in Tables 7 and 8 .

Table 7: Alarm Message Priority

Table 8: Alarm Tone Priority

Priority

Message

Priority

Tone

1 (Highest) Alarm Message

1 (Highest) Button Acknowledge Tone

2

Advisory Message

2

Invalid Tone

3

Informational Message

3

Alarm Tone

4

RF Stop Tone

5

Informational Tone

6

RF Start Tone

7

RF On Tone

8

RF Pullback Pause Tone

9

Power On Tone

Summary of Contents for VNUS RFG2

Page 1: ...VNUS Radio Frequency Generator RFGPLUS Model RFG2 Service Manual...

Page 2: ...iance 8 Biomedical Department Power OutputTest 12 Power DeliveryVerification 13 Alarm Priority 13 Error Codes 14 Service and Maintenance 16 Cleaning and Disinfection 16 Repairs and Returns 16 Warranty...

Page 3: ...ator The RF Switch circuitry contained on the UI Board provides the switch and colored backlighting for the RF on off switch located on the right of the FPI as viewed from the front of the Generator R...

Page 4: ...nly be removed with a tool Service Port is UL1950 50VDC Normal MAINSV in Single Fault Condition Metal Enclosure is Protective Earth Display has no touch screen and sits behind protective plastic cover...

Page 5: ...TL100 119 02 1 Set the line voltage to 264V and set the frequency to 50Hz DO NOT EXCEED 280V 2 Plug the line cord of theVNUS Generator into the leakage tester for all tests 3 For source leakage and i...

Page 6: ...tion Nearby electronic equipment may be affected RF emissions CISPR 11 Class A The RFG2 is suitable for use in all establishments other than domestic and those directly connected to the public low vol...

Page 7: ...e considered If the measured field strength in the location in which the RF Generator is used exceeds the applicable RF compliance level above the RF Generator should be observed to verify normal oper...

Page 8: ...nce thermometer and the ESU analyzer Turn off main power to the generator Remove the modified Instrument Cable from the RF Generator and AC power cord from the generator In assessing the functionality...

Page 9: ...es Turn off the unit wait a few seconds and then turn on the unit If the error persists try a different MMC Error 26 The inserted Multimedia Card MMC cannot upgrade this RF Generator because its seria...

Page 10: ...tative will assist the operator with obtaining service or returning the RF Generator as appropriate The RF Generator must be unplugged before servicing WARNING Do not remove the cover of the RF Genera...

Page 11: ...e SJT16 3 IEC color code maximum length 10 ft 3 m Plug minimum 10 A 125VAC Unit receptacle IEC female minimum 10 A 125VAC 220 240VAC Cable H05VVF3G1 0VDE maximum length 10 ft 3 m Plug minimum 6 A 125V...

Page 12: ...0 10 30 40 50 60 70 80 90 100 0 5 10 15 20 25 30 35 40 Peak Voltage V Power W Figure 8 Peak Output Voltage vs Power Output 0 0 5 5 10 10 15 15 20 20 25 30 35 40 25 30 35 40 Power Output W Max Power Se...

Page 13: ...88 797 VEIN 8346 Fax 1 408 365 8470 Covidien Ireland Limited IDA Business Technology Park Tullamore 0123 Customer Service Tel 1 800 962 9888 MEDICAL EQUIPMENT MODEL RFG2 WITH RESPECT TO ELECTRICAL SHO...

Reviews: