background image

Kendall SCD

 

 

700 Sequential Compression System

Operation and Service Manual

700 Système de compression séquentielle

 

- Manuel d’utilisation et d’entretien

700 Sequenzielles Kompressionssystem

 

- Bedienungs- und Servicehandbuch

Sistema di compressione sequenziale 700

 

- Manuale d’uso e di manutenzione

Sistema de compresión secuencial 700

 

- Manual de funcionamiento y mantenimiento

700 Sekventiellt kompressionssystem

 

- Användar- och servicehandbok

700 sequentieel compressiesysteem

 

- Bedienings- en onderhoudshandleiding

Sistema de Compressão Sequencial 700

 

- Manual de Funcionamento e Assistência

700 jaksoittainen kompressiojärjestelmä

 

- Käyttö- ja huolto-ohjekirja

700 Sekventielt kompressionssystem

 

- Bruger- og servicevejledning

Σύστημα διαδοχικής συμπίεσης 700

 

- Εγχειρίδιο λειτουργίας και σέρβις

Sekvenční kompresní systém 700

 

- Uživatelská a servisní příručka

700 Szekvenciális kompressziós rendszer

 

- Kezelési és szervizelési kézikönyv

700 Система терапевтическая для  

 

 

последовательной компрессии

 

- Руководство по эксплуатации и  

 

 

 

 

  обслуживанию

System stopniowanego ucisku 700 

 

- Podręcznik obsługi i serwisu

700 Sıralı Kompresyon Sistemi

 

- Çalıştırma ve Servis El Kitabı

700 Sekvensielt kompresjonssystem

 

- Bruker- og servicehåndbok

Sekvenčný kompresný systém 700

 

- Príručka na obsluhu a servis

Sistem de compresie secvenţială 700

 

- Manual de operare şi întreţinere

Система за последователна компресия 700

 

- Ръководство за работа и сервиз

Summary of Contents for Kendall SCD 700 Series

Page 1: ...gs en onderhoudshandleiding Sistema de Compress o Sequencial 700 Manual de Funcionamento e Assist ncia 700 jaksoittainen kompressioj rjestelm K ytt ja huolto ohjekirja 700 Sekventielt kompressionssyst...

Page 2: ...it Powered Off charging when plugged in EN 8 Charging the Battery EN 8 BatteryWarnings EN 8 Section III Fault Conditions andTroubleshooting EN 9 Section IV Service and Maintenance EN 13 Introduction E...

Page 3: ...an Filter Removal Installation see Figure 6 EN 20 Main CPU Board and Graphical Display Removal Installation see Figure 6 EN 20 Adjustable Bed Hook Removal Installation EN 20 SectionVII Parts Listing E...

Page 4: ...tion enhancement 2 Deep vein thrombosis prophylaxis 3 Edema Acute 4 Edema Chronic 5 Extremity pain incident to trauma or surgery 6 Leg Ulcers 7 Venous stasis venous insufficiency If you need further i...

Page 5: ...e power cord is damaged WARNING Donotattempttorepairorreplacebrokentubingconnectorsashazardousinflationofthesleevesmayoccur Kendall SCD 700 Series Explanation of Symbols Used Caution consult accompany...

Page 6: ...mity to the point of use Be sure to allow adequate air flow to the vents located at the power cord cover and below the tube set connection points The controller can operate with one or two garments at...

Page 7: ...ed The A and B buttons need to be pressed only when foot compression is to be used NOTE If a garment is attached anytime after the Garment Detection procedure has started the system must be restarted...

Page 8: ...e process of applying intermittent compression alternating between limbs or to one if only one garment is applied On successive cycles the controller automatically adjusts its operating parameters to...

Page 9: ...high Length Medium 73013 Thigh Length Large 73022 Knee Length Medium 73023 Knee Length Large Kendall SCD Sequential Compression ComfortTear Away Sleeves 74041 Thigh Length Small 74042 Thigh Length Med...

Page 10: ...GURE 3 Warning If the ground integrity of the mains power cable is in question the device should be operated on battery power until the ground integrity can be insured FIGURE 3 1 3 2 Battery Status In...

Page 11: ...epending on the sleeve configuration sleeve application and the battery condition Note If the operation time on battery power is extremely short the battery should be returned for service or replaceme...

Page 12: ...s Description Example Service Required Alarm code is present because of a failed internal component It can not be addressed by the user E5 Manual Reset Required Alarm that can be troubleshot and corre...

Page 13: ...User Resettable Leg Sleeve pressure is greater than 47 mmHg for 10 consecutive cycles or pressure is above 65 mmHg for 5 consecutive cycles Check for a tight leg sleeve and adjust fit appropriately A...

Page 14: ...ired If a valve electrically malfunctions this error will be displayed ServiceTechnician only Verify that the valve assembly wires are properly connected and confirm solenoid actuation E5 Software Err...

Page 15: ...been made Replace pack or return to Covidien for service E10 Tubing Disconnect Alarm User Resettable Pressure measured in the inflatable garment is below 10 mmHg for 10 consecutive cycles or no garmen...

Page 16: ...one year of delivery to the original purchaser Specifically we agree to service and or adjust any Controller as required if returned for that purpose and to replace and repair any part which upon our...

Page 17: ...tential lug located on the back of the device opposite the power cord There are no other grounded exposed metal parts Power cord resistance should not exceed 0 1 ohm If ground resistance exceeds this...

Page 18: ...nted circuit boards In manufacturing extensive testing is performed that cannot be duplicated in the field without specialized equipment Improper repair could result in patient or user hazards Pneumat...

Page 19: ...eT1 The buzzer will not pip during this alarm Test ModeT2 General FunctionTest With nothing plugged into the ports on the back of the controller enter test access mode SelectTest Access ModeT2 Press t...

Page 20: ...onthebackofthecontroller entertestaccessmode SelectTestAccessModeT4 Press the B Button to begin the test TheT4 will blink on the display screen until the calibration verification procedure is complete...

Page 21: ...ible alarm After the B Button is pressed the B Leg icon will stop flashing the alarm will stop and the controller will then go through a normal inflation cycle at B with the pump operating at a high s...

Page 22: ...Installation see Figure 8 The compressor is not a user serviceable component Do not disassemble Do not oil The compressor is held in place by friction of its molded foam enclosure Disconnect the compr...

Page 23: ...to pull air through the power cord door Note the molded arrow in the fan case showing the flow direction For optimum cooling and quietness use only Covidien replacement fans Main CPU Board and Graphi...

Page 24: ...d hook Assembly Main CPU printed circuit board Power supply circuit board LCD Display Membrane switch panel Power cord Power cord UK Power cord Europe Power cord Japan Power cord Australia New Zealand...

Page 25: ...Power Cord Storage Yes Audible Visual Alarms Low Pressure High Pressure Internal Electronics Malfunction Power Cord 13 feet long with region specific appropriate cordage and plug Controller Dimension...

Page 26: ...SCD700Seriesshould assurethatitisusedinsuchanenvironment Immunity test IEC 60601 Test level Compliance level Electromagnetic environment guidance Electrostaticdischarge ESD IEC61000 4 2 6kVcontact 8k...

Page 27: ...cy Toassesstheelectromagneticenvironmentduetofixed RFtransmitters anelectromagneticsitesurveyshouldbeconsidered IfthemeasuredfieldstrengthinthelocationinwhichtheSCD700Series controllerisusedexceedsthe...

Page 28: ...3 Power supply circuit board 4 Valve manifold assembly 5 Fuse pair 6 Battery pack 7 Fan assembly 8 USB connector 9 Muffler 10 Rear Enclosure assembly 11 Power cord see spare parts insert for country s...

Page 29: ...EN 26 Kendall SCD 700 Series 19 20 16 1 17 18 Figure 6 Parts Assembly Diagram front enclosure Exploded view Page 2 of 2...

Page 30: ...J9 CPU TO J2 CPU J2 POWER SUPPLY BOARD FAN J1 EQUIPOTENTIAL LUG MAINS VOLTAGE INPUT 100 240 VAC 50 60 Hz J9 J3 POWER SUPPLY J4 AC INLET FUSES J6 J5 J1 TO J4 SUPPLY POWER FLOW DIRECTION CPU TO J6 COMPR...

Page 31: ...EN 28 Kendall SCD 700 Series Figure 8 Rear EnclosureView Muffler Equipotential Lug Location Compressor Assembly Power Supply Battery...

Page 32: ...EN 29 Kendall SCD 700 Series Figure 9 Front EnclosureView Main CPU Board...

Page 33: ...ida o que no est presente dentro del producto o embalaje Identifiering av ett mne som inte ing r eller f rekommer i produkten eller f rpackningen Identificatie van een stof die niet in het product of...

Reviews: