28 - Installation - Quark Installation
Quark Unit Rear Panel
Connection Description
Quark
Model
1. PNT/MIP-MEP/P0.1 port
Pneumatic connection for X9 PNT and MIP-MEP/P0.1
PFT, SPIRO
2. Permapure sampling line port
Connection for the sample line, luer lock, rotate clockwise to tighten
PFT, CPET
3. Breathing alve connector
Large connector for Breathing Valve Assembly
PFT
4. DLCO cylinder connector
Push connector with quick release for DLCO gas cylinder
PFT
5. O
2
/CO
2
calibration cylinder connector
Push connector with quick release for O
2
/CO
2
calibration gas cylinder
PFT, CPET
6. Oxygen cylinder connector
Push connector with quick release for O
2
gas cylinder
PFT
7. Demand valve pneumatic connection
Connection for demand valve for gas delivery from S/N 2014041001
PFT (newer)
8. Supplemental ground (earthing)
Connection to ground the unit
PFT, SPIRO, CPET
9. Dosimeter Port
Connections for controlling delivery of medications to subject
PFT, SPIRO
10. Medical air connector
Push connector with quick release for medical air to control dosimeter
11. RS232 port for PC
Serial connection to computer (COM Port)
PFT, SPIRO, CPET
12. Temperature / Relative Humidity Port
Connection for Temperature and Humidity Probe (external)
PFT, SPIRO, CPET
13. USB-A port to PC
USB connection to computer (printer style cable)
PFT, SPIRO, CPET
14. Auxiliary USB port for HR Monitor
USB connection (USB Drive connection) for Heart Rate Monitor Dongle
PFT, CPET
15. Turbine flowmeter port
Phone style connection for connection of Turbine Flowmeter(s)
PFT, SPIRO, CPET
16. Reset button
Button to reset system, power button is preferred method
PFT, SPIRO, CPET
17. Auxiliary RS232 port – Ergometers
Communication with Ergometers / accessories via RS232
PFT, CPET
18. Oximeter/PNT ROCC connector (for PNT ROCC, an
adapter is required)
Connection for Oximeter(s) and the PNT Rocc
PFT, CPET
19. Auxiliary TTL input/output port
Auxiliary connection for input / output of TTL signals (ECG / Polar HRM)
PFT, CPET
20. Port for the soda lime CO
2
absorber output (pre-
installed)
Luer lock connection for soda lime absorber, rotate clockwise to tighten
PFT, CPET
21. Power cable plug
Plug to insert appropriate power cord into
PFT, SPIRO, CPET
22. Soda Lime CO
2
Absorber slot (pre-installed)
Carbon Dioxide absorber holder
PFT, CPET
1.
1
2
3
4 5 6
7
8 9
10
11 12
13 15
14 16
17
18 19
20
21
22
Summary of Contents for Quark CPET
Page 2: ......
Page 8: ......
Page 9: ...Getting Started...
Page 25: ...Installation...
Page 43: ...System Cleaning Maintenance...
Page 48: ...46 System Cleaning Maintenance Spirometry Tests Spirometry Tests...
Page 56: ...54 System Cleaning Maintenance CPET RMR Tests CPET RMR Tests...
Page 63: ...System Cleaning Maintenance Canopy test 61 Canopy test...
Page 68: ...66 System Cleaning Maintenance Mixing Chamber Test Mixing Chamber Test...
Page 77: ...System Cleaning Maintenance Rocc test 75 Rocc test...
Page 83: ...Appendix...
Page 101: ......