22 - Getting Started - System Warm-Up
System Warm-Up
Each Quark unit must be warmed up prior to use for the required amount of time. The warm-up time depends on which test is being performed.
Below is a table displaying minimum warm-up times based on each test:
Test Warm-up
Time
(mins)
FVC, VC, MVV
-
FRC
5
CV
5
DLCO
15
MIP/MEP
-
P0.1
-
Oximetry
-
R
OCC
-
CPET
5
RMR
5
During the warm-up, the device must be powered on, however the software does not need to be open.
Calibration and/or testing procedures should not be performed until the warm-up period has been completed.