Instruction Manual   |   7

1

2

3

4

5

6

Rest position

First stop

Second stop

Aspiration

1. Press the pushbutton to the first positive stop. Holding 

the pipettor vertically, immerse the pipet tip into the 

sample liquid. The depth to which the pipet tip is 

immersed in the sample liquid depends on the model.

Model Volume Range (μL)

Immersion Depth (mm)

2, 10

≤1

20, 50, 100

2 - 3

200, 1,000

2 - 4

2. Release the pipetting pushbutton slowly and smoothly 

to aspirate the sample. Wait one second and then 

withdraw the pipet tip from the liquid. If the pipet tip 

is not immersed to the recommended depth or if the 

pipetting pushbutton is rapidly released, air may enter 

the disposable pipet tip.

NOTE:

 Avoid touching the orifice of the pipet tip.

Dispensing

3. Place the end of the pipet tip against the inside wall 

of the vessel at an angle of 10° to 40°. Press the 

pushbutton smoothly to the first stop. Wait one second. 

4. Press the pushbutton to the second stop to expel any 

remaining liquid. Keep the pushbutton depressed to the 

very end, and remove the pipettor by drawing the pipet 

tip against the inside surface of the receiving vessel.

5. Release the pushbutton to its starting position.
6. Eject the pipet tip by pressing the tip ejector button. 

Remember to change the pipet tip whenever a different 

type of liquid is to be sampled.

Summary of Contents for Lambda Plus

Page 1: ...Corning Lambda Plus Pipettors Single channel 8 channel and 12 channel Pipettors Instruction Manual...

Page 2: ...ng the Volume 5 4 Aspiration and Dispensing Instructions 6 5 Pre rinsing 8 6 Dense and Viscous Liquids 8 7 Corning Pipet Tips 9 8 Recommendations 9 9 Recalibration 10 10 Troubleshooting 13 11 Cleaning...

Page 3: ...ettors are produced in four volume ranges 0 5 to 10 L 5 to 50 L 20 to 200 L 50 to 300 L Model Volume Range L Application 2 10 Measurement and transfer of micro volumes DNA sequencing and enzyme assay...

Page 4: ...jector D Single channel Pipettor Ejector The ejector length can be adjusted to accommodate all common brands of pipet tips When using narrow tubes it may be necessary to remove the tip ejector This is...

Page 5: ...ng at least 10 measurements of distilled water at a reference temperature of 20 C 1 C according to EN ISO 8655 standard 8 channel and 12 channel Pipettor Ejector The 8 and 12 channel pipettor construc...

Page 6: ...the test volume selected Precision is the degree to which repeated measurements under unchanged conditions show the same results The accuracy and precision are obtained gravimetrically using Corning...

Page 7: ...nd red Red numbers represent decimal fractions for microvolume pipettors or milliliters for macrovolume pipettors An example for each of these pipettors is given on the next page Nominal Volume L Coun...

Page 8: ...too far below the required value the process must be repeated Set the locking ring into the upper position following the volume adjustment thus locking the knob and preventing accidental volume chang...

Page 9: ...to the recommended depth or if the pipetting pushbutton is rapidly released air may enter the disposable pipet tip NOTE Avoid touching the orifice of the pipet tip Dispensing 3 Place the end of the pi...

Page 10: ...sion are based on pipetting distilled water Handling of liquids with physical qualities of density viscosity and surface tension which differ substantially from water may need to be gravimetrically ch...

Page 11: ...ry and should remain constant during aspiration The pipettor should be held in a vertical position Change the pipet tip when volume setting is changed or when a different liquid is to be aspirated Cha...

Page 12: ...ical Density of the liquid used is close to that of distilled water Balance with appropriate sensitivity has been used Conversion factor mg L has been taken into account The requirements shown in Sect...

Page 13: ...0 33 8 50 12 50 5 50 5 4 8 5 2 1 67 8 200 12 200 20 200 20 19 4 20 6 6 30 8 300 12 300 50 300 50 49 2 50 8 10 00 Perform 5 aspirations weigh each one and calculate the average value of the aspiration...

Page 14: ...o reduce the aspirated volume or counterclockwise to increase the volume One full turn of the calibration key changes the pipettor aspiration volume by the amount given in the table 4 Take out the cal...

Page 15: ...ns The pipet tip is incorrectly pressed onto the pipettor shaft Press the pipettor firmly The tip is damaged or worn out due to extensive use Replace the tip with a new one The pipettor incorrectly as...

Page 16: ...eps No lubricant on the sealing elements Remove the tip ejector Unscrew the shaft nut and remove the shaft piston assembly seal and O ring Rinse the removed parts in distilled water and dry thoroughly...

Page 17: ...ried and cooled to room temperature It is recommended to Unscrew the shaft nut slightly prior to sterilization and re tighten after autoclaving Set the locking ring in lower unlocked position prior to...

Page 18: ...ion key M Ejector cap N Ejector pushbutton P Identification ring S Locking ring Please contact Corning Customer Service to inquire about availability of spare parts Pipettor model and name of the part...

Page 19: ...not arising from defects in original material or workmanship This warranty does not cover the PTFE seal O ring or shaft Claims for transit damage should be filed with the transportation carrier In the...

Page 20: ...re for research use or general laboratory use only Not intended for use in diagnostic or therapeutic procedures Not for use in humans These products are not intended to mitigate the presence of microo...

Page 21: ...wan t 886 2 2716 0338 EUROPE CSEurope corning com France t 0800 916 882 Germany t 0800 101 1153 The Netherlands t 20 655 79 28 United Kingdom t 0800 376 8660 All Other European Countries t 31 0 206 59...

Reviews: