background image

14   |   CORNING LAMBDA PLUS PIPETTORS 

Problem

Cause

Solution

The pipettor 

incorrectly 

aspirates 

the liquid or 

liquid drops 

out from the 

pipet tip.

Damage to the 

piston or seal due 

to prolonged use 

with chemically 

aggressive liquids.

Replace the piston, seal, and 

O-ring. Rinse the inside of 

the shaft in distilled water 

and dry. Lubricate the seal 

and O-ring with the lubricant 

which has been included 

with each pipettor. The 

replacement of the piston 

requires conducting the 

calibration procedure.

NOTE:

 The parts of the single-

channel (2 μL and 10 μL) 

pipettors should be lubricated 

evenly with a minimum 

amount of lubricant.

The pipettor is 

reassembled 

improperly.

Disassemble the pipettor and 

reassemble it, observing the 

proper sequence of steps.

No lubricant on the 

sealing elements.

Remove the tip ejector. 

Unscrew the shaft nut and 

remove the shaft, piston 

assembly, seal, and O-ring. 

Rinse the removed parts 

in distilled water and dry 

thoroughly. Lightly lubricate 

the inside surfaces of the 

seal and the O-ring with 

the included lubricant. 

Reassemble the pipettor in 

the reverse order.

Contamination of 

the inside of the 

pipettor caused by 

extensive aspiration 

of chemically 

aggressive liquids or 

because liquid got 

inside the pipettor.

Contamination of the inside 

of the pipettor caused by 

extensive aspiration of 

chemically aggressive liquids 

or because liquid got inside 

the pipettor.

If the problem continues after carrying out the above 

steps, contact your Corning representative.
Before returning the pipettor, please ensure the pipettor 

is completely free of any chemical, radioactive, or 

microbiological contamination which could pose a threat 

during transport and repair.

Summary of Contents for Lambda Plus

Page 1: ...Corning Lambda Plus Pipettors Single channel 8 channel and 12 channel Pipettors Instruction Manual...

Page 2: ...ng the Volume 5 4 Aspiration and Dispensing Instructions 6 5 Pre rinsing 8 6 Dense and Viscous Liquids 8 7 Corning Pipet Tips 9 8 Recommendations 9 9 Recalibration 10 10 Troubleshooting 13 11 Cleaning...

Page 3: ...ettors are produced in four volume ranges 0 5 to 10 L 5 to 50 L 20 to 200 L 50 to 300 L Model Volume Range L Application 2 10 Measurement and transfer of micro volumes DNA sequencing and enzyme assay...

Page 4: ...jector D Single channel Pipettor Ejector The ejector length can be adjusted to accommodate all common brands of pipet tips When using narrow tubes it may be necessary to remove the tip ejector This is...

Page 5: ...ng at least 10 measurements of distilled water at a reference temperature of 20 C 1 C according to EN ISO 8655 standard 8 channel and 12 channel Pipettor Ejector The 8 and 12 channel pipettor construc...

Page 6: ...the test volume selected Precision is the degree to which repeated measurements under unchanged conditions show the same results The accuracy and precision are obtained gravimetrically using Corning...

Page 7: ...nd red Red numbers represent decimal fractions for microvolume pipettors or milliliters for macrovolume pipettors An example for each of these pipettors is given on the next page Nominal Volume L Coun...

Page 8: ...too far below the required value the process must be repeated Set the locking ring into the upper position following the volume adjustment thus locking the knob and preventing accidental volume chang...

Page 9: ...to the recommended depth or if the pipetting pushbutton is rapidly released air may enter the disposable pipet tip NOTE Avoid touching the orifice of the pipet tip Dispensing 3 Place the end of the pi...

Page 10: ...sion are based on pipetting distilled water Handling of liquids with physical qualities of density viscosity and surface tension which differ substantially from water may need to be gravimetrically ch...

Page 11: ...ry and should remain constant during aspiration The pipettor should be held in a vertical position Change the pipet tip when volume setting is changed or when a different liquid is to be aspirated Cha...

Page 12: ...ical Density of the liquid used is close to that of distilled water Balance with appropriate sensitivity has been used Conversion factor mg L has been taken into account The requirements shown in Sect...

Page 13: ...0 33 8 50 12 50 5 50 5 4 8 5 2 1 67 8 200 12 200 20 200 20 19 4 20 6 6 30 8 300 12 300 50 300 50 49 2 50 8 10 00 Perform 5 aspirations weigh each one and calculate the average value of the aspiration...

Page 14: ...o reduce the aspirated volume or counterclockwise to increase the volume One full turn of the calibration key changes the pipettor aspiration volume by the amount given in the table 4 Take out the cal...

Page 15: ...ns The pipet tip is incorrectly pressed onto the pipettor shaft Press the pipettor firmly The tip is damaged or worn out due to extensive use Replace the tip with a new one The pipettor incorrectly as...

Page 16: ...eps No lubricant on the sealing elements Remove the tip ejector Unscrew the shaft nut and remove the shaft piston assembly seal and O ring Rinse the removed parts in distilled water and dry thoroughly...

Page 17: ...ried and cooled to room temperature It is recommended to Unscrew the shaft nut slightly prior to sterilization and re tighten after autoclaving Set the locking ring in lower unlocked position prior to...

Page 18: ...ion key M Ejector cap N Ejector pushbutton P Identification ring S Locking ring Please contact Corning Customer Service to inquire about availability of spare parts Pipettor model and name of the part...

Page 19: ...not arising from defects in original material or workmanship This warranty does not cover the PTFE seal O ring or shaft Claims for transit damage should be filed with the transportation carrier In the...

Page 20: ...re for research use or general laboratory use only Not intended for use in diagnostic or therapeutic procedures Not for use in humans These products are not intended to mitigate the presence of microo...

Page 21: ...wan t 886 2 2716 0338 EUROPE CSEurope corning com France t 0800 916 882 Germany t 0800 101 1153 The Netherlands t 20 655 79 28 United Kingdom t 0800 376 8660 All Other European Countries t 31 0 206 59...

Reviews: