background image

8   |   CORNING LAMBDA PLUS PIPETTORS 

 5.  Pre-rinsing

When pipetting liquids of higher viscosity or lower surface 

tension than water (e.g., sera or organic solvents), a film 

of liquid is formed on the inside wall of the pipet tip. This 

film can create an error. As the film remains relatively 

constant in successive pipetting operations with the same 

tip, this error can be eliminated by forming the film before 

transferring the first sample. This is done by aspirating a 

sample and dispensing it back into the same vessel. Now, 

as the film is already formed, all of the following samples 

will have better accuracy and repeatability.
This pre-rinsing operation should be repeated when the 

volume to be aspirated is changed or when a new pipet tip 

is used.

 6. Dense and Viscous Liquids

The pipettor’s specifications of accuracy and precision are 

based on pipetting distilled water. Handling of liquids with 

physical qualities of density, viscosity, and surface tension 

which differ substantially from water may need to be 

gravimetrically checked for compensation of the volume 

setting. Normally the degree of error resulting from heavy 

or viscous liquids is negligible if pipetting is done slowly 

and carefully. It is most important to give the liquids 

some time to react to the change of pressure by holding 

the pipet tip in its position for at least 2 seconds after the 

aspiration and the blow out stroke.
If in extreme cases, this method of operation does not 

result in accurate values, compensation may be achieved 

as follows: 

 

Weigh the pipetted liquid when the pipettor is set to the 

nominal value. Then calculate the set-off from the 

nominal value: 

        Correction, value = 2 x nominal val. —

        m = weight of the sample
        

γ

 = density of liquid

 

Check this operation once again and correct if necessary.

Note the corrected value for further pipetting of the 

same kind of liquid.

γ

m

Summary of Contents for Lambda Plus

Page 1: ...Corning Lambda Plus Pipettors Single channel 8 channel and 12 channel Pipettors Instruction Manual...

Page 2: ...ng the Volume 5 4 Aspiration and Dispensing Instructions 6 5 Pre rinsing 8 6 Dense and Viscous Liquids 8 7 Corning Pipet Tips 9 8 Recommendations 9 9 Recalibration 10 10 Troubleshooting 13 11 Cleaning...

Page 3: ...ettors are produced in four volume ranges 0 5 to 10 L 5 to 50 L 20 to 200 L 50 to 300 L Model Volume Range L Application 2 10 Measurement and transfer of micro volumes DNA sequencing and enzyme assay...

Page 4: ...jector D Single channel Pipettor Ejector The ejector length can be adjusted to accommodate all common brands of pipet tips When using narrow tubes it may be necessary to remove the tip ejector This is...

Page 5: ...ng at least 10 measurements of distilled water at a reference temperature of 20 C 1 C according to EN ISO 8655 standard 8 channel and 12 channel Pipettor Ejector The 8 and 12 channel pipettor construc...

Page 6: ...the test volume selected Precision is the degree to which repeated measurements under unchanged conditions show the same results The accuracy and precision are obtained gravimetrically using Corning...

Page 7: ...nd red Red numbers represent decimal fractions for microvolume pipettors or milliliters for macrovolume pipettors An example for each of these pipettors is given on the next page Nominal Volume L Coun...

Page 8: ...too far below the required value the process must be repeated Set the locking ring into the upper position following the volume adjustment thus locking the knob and preventing accidental volume chang...

Page 9: ...to the recommended depth or if the pipetting pushbutton is rapidly released air may enter the disposable pipet tip NOTE Avoid touching the orifice of the pipet tip Dispensing 3 Place the end of the pi...

Page 10: ...sion are based on pipetting distilled water Handling of liquids with physical qualities of density viscosity and surface tension which differ substantially from water may need to be gravimetrically ch...

Page 11: ...ry and should remain constant during aspiration The pipettor should be held in a vertical position Change the pipet tip when volume setting is changed or when a different liquid is to be aspirated Cha...

Page 12: ...ical Density of the liquid used is close to that of distilled water Balance with appropriate sensitivity has been used Conversion factor mg L has been taken into account The requirements shown in Sect...

Page 13: ...0 33 8 50 12 50 5 50 5 4 8 5 2 1 67 8 200 12 200 20 200 20 19 4 20 6 6 30 8 300 12 300 50 300 50 49 2 50 8 10 00 Perform 5 aspirations weigh each one and calculate the average value of the aspiration...

Page 14: ...o reduce the aspirated volume or counterclockwise to increase the volume One full turn of the calibration key changes the pipettor aspiration volume by the amount given in the table 4 Take out the cal...

Page 15: ...ns The pipet tip is incorrectly pressed onto the pipettor shaft Press the pipettor firmly The tip is damaged or worn out due to extensive use Replace the tip with a new one The pipettor incorrectly as...

Page 16: ...eps No lubricant on the sealing elements Remove the tip ejector Unscrew the shaft nut and remove the shaft piston assembly seal and O ring Rinse the removed parts in distilled water and dry thoroughly...

Page 17: ...ried and cooled to room temperature It is recommended to Unscrew the shaft nut slightly prior to sterilization and re tighten after autoclaving Set the locking ring in lower unlocked position prior to...

Page 18: ...ion key M Ejector cap N Ejector pushbutton P Identification ring S Locking ring Please contact Corning Customer Service to inquire about availability of spare parts Pipettor model and name of the part...

Page 19: ...not arising from defects in original material or workmanship This warranty does not cover the PTFE seal O ring or shaft Claims for transit damage should be filed with the transportation carrier In the...

Page 20: ...re for research use or general laboratory use only Not intended for use in diagnostic or therapeutic procedures Not for use in humans These products are not intended to mitigate the presence of microo...

Page 21: ...wan t 886 2 2716 0338 EUROPE CSEurope corning com France t 0800 916 882 Germany t 0800 101 1153 The Netherlands t 20 655 79 28 United Kingdom t 0800 376 8660 All Other European Countries t 31 0 206 59...

Reviews: