![Coremorrow S51.ZT2K-C1 User Manual Download Page 12](http://html1.mh-extra.com/html/coremorrow/s51-zt2k-c1/s51-zt2k-c1_user-manual_2667312012.webp)
S51.ZT2S/K-C1 Piezo Z/Tip/Tilt Platform User Manual
10
5.3 Operating
Please reconfirm the following information before turning on the power:
S51.ZT2S/K-C1 has been grounded.
The operating environment meets the operating environment requirements of it.
Fixed screws must be tightened.
After the above information is confirmed, relevant controller operations can be carried out:
1.Please read the user manual of the controller carefully before operation, and then start
the controller.
2.Set the controller' output voltage to zero first.
3.Check no abnormal phenomenon, then input voltage value (via keyboard or analog
control) according to specified input voltage range.
4.Operating normally within the rated voltage range of piezoelectric ceramics (0~120V,
maximum 150V).
5.After use, please carry out voltage reset operation.
Please take light steps when S51.ZT2S/K-C1 in operation, do not knock the product, so as
not to affect the use of the product or even damage the product.
In order to achieve better motion precision of S51.ZT2S/K-C1, it is recommended to use
our company's supporting controller. Please refer to the user manual of the controller for
the specific operation to ensure correct operation before relevant operation.
Notice:
Do not remove the ground protection during operation. If it needs to be temporarily
removed, please reconnect the ground protection before restart.
The driving voltage shall not exceed the voltage range of S51.ZT2S/K-C1.
If any abnormal noise or oscillation occurs during use, turn off the power immediately
and check the parameter setting.