3
In order to prevent damage to the product caused by improper operation, the
following points should be paid attention to when in use.
P78 Piezo Z Microscope Scanner is used in a dust-free, oil-free and lubricant-free
environment.
As the product is designed with flexible structure, it is recommended that the load
should not exceed load capacity of product. When loading, DO NOT apply torque
force and DO NOT pry or twist the load surface, so as not to damage the structure.
Recommended voltage range is 0~120V for long-term use and lifetime.
Do not disassemble the product to avoid damage.
Avoid stretching and bending the cable interface to prevent damage to the cable.
Use the dedicated cable provided by CoreMorrow to connect P78 Piezo Z
Microscope Scanner to controller.
Do not connect via an extension cable casually, if you need a longer cable, please
contact CoreMorrow.
A change in temperature and pressure will charge the piezoelectric actuator, and
the actuator will remain energized for a period of time when it is disconnected
from the controller, So the operator should pay attention to when in use:
Do not disassemble P78 Piezo Z Microscope Scanner.
Discharge P78 before installation, which can be done by connecting to the
controller.
Do not unplug the wire connecting with controller during operation.
Before use, check whether the connection wire of the product is in good condition and
whether the product and the controller are effectively grounded. During operation,
strictly standardize the operation. Do not touch the product (up to 150V voltage) with
your hand when power on, in case of danger.
2. Safety Guide
P78 Piezo Z Microscope Scanner is designed with the most advanced technology and
safety standards for your own safety and the correct use of the product, please observe the
following points when in use.