P78 Piezo Z Microscope Scanner User Manual
6
4. Operating environment
A clean, dustfree and non-corrosive environment
Operating temperature : -20~80℃
Relative humidity : <60%
Recommended voltage range is 0~120V for long-term use and lifetime
5. Installation and operation
Safety check should be carried out before installation, such as whether the connection wire
is damaged or broken, whether the drive power supply is grounded, etc.
5.1 Installation and Precautions
Installation diagram of P78 is as follows:
FIG.2 P78 Installation diagram
(b) P78.Z200/300S/K
(a) P78.Z100S/K
Load mounting(M3)
Hole for grounding
(M3)
Motion surface
Mounting(M4)
Load mounting(M2.5)
Mounting(
φ
6.4THRU)