After the above information is confirmed, relevant controller operations can be carried out:

1.Please read the user manual of the controller carefully before operation, and then start 

the controller.

2.Set the controller' output voltage to zero first.

3.Check no abnormal phenomenon, then input voltage value (via keyboard or analog 

control) according to specified input voltage range.

4.Operating normally within the rated voltage range of piezoelectric ceramics (0~120V, 

maximum 150V).

5.After use, please carry out voltage reset operation.

Please take light steps when P78 in operation, do not knock the product, so as not to 

affect the use of the product or even damage the product.

In order to achieve better motion precision of P78, it is recommended to use our 

company's supporting controller. Please refer to the user manual of the controller for the 

specific operation to ensure correct operation before relevant operation.

Notes: 

Do not remove the ground protection during operation. If it needs to be temporarily 

removed, please reconnect the ground protection before restart. 

The driving voltage shall not exceed the voltage range of P78.

If any abnormal noise or oscillation occurs during use, turn off the power immediately 

and check the parameter setting

6.1 Transportation

This product is packed in cartons. Transportation is carried out under the packaging 

conditions. This product can be transported by various means of transport under normal 

circumstances. In the process of transportation, direct rain and snow, corrosive substances, 

6. Transportation and inspection

Summary of Contents for P78 Piezo Z

Page 1: ...P78 Piezo Z Microscope Scanner User Manual Version V1 0 This document describes the following products P78 Z100S P78 Z200S K P78 Z300S K...

Page 2: ...you disassemble or transform the product without permission our company will not be responsible for any consequences arising therefrom To change the product model and other false sales of our company...

Page 3: ...5 4 Operating environment 6 5 Installation and operation 6 5 1 Installation and Precautions 6 5 2 Load 8 5 3 Operating 10 6 Transportation and inspection 11 6 1 Transportation 11 6 2 Open Box Inspecti...

Page 4: ...and use 1 2 Symbol and meaning DANGER Improper handling may could cause injury to the operator Attention Improper handling can cause damage to the equipment No order between items 1 3 Users of This Ma...

Page 5: ...casually if you need a longer cable please contact CoreMorrow A change in temperature and pressure will charge the piezoelectric actuator and the actuator will remain energized for a period of time wh...

Page 6: ...e structure P78 Piezo Z Microscope Scanner adopting flexible hinge mechanism has advantages of no friction high positioning accuracy Flexible hinge is an element not affected by static and dynamic fri...

Page 7: ...f up to 300 m in Z and features fast response speed no mechanical friction high repeatable positioning accuracy easy to operate so it can meet different applications This product can also be customize...

Page 8: ...e 5 Installation and operation Safety check should be carried out before installation such as whether the connection wire is damaged or broken whether the drive power supply is grounded etc 5 1 Instal...

Page 9: ...and tighten it properly to ensure the positioning accuracy 5 Connect the P78 connector with the corresponding controller according to the identification The wiring mode of Lemo connector is shown as...

Page 10: ...osition to mount the load is shown below FIG 5 The load is applied to the center of the moving surface FIG 6 Try not to let the position of the center of gravity far exceed the active platform For the...

Page 11: ...in the fastening and connecting process will lead to failure in installation or leakage of bolts which will expose the product and scratch the product affecting the motion accuracy of the product Sho...

Page 12: ...echanism Avoid stretching and bending cables Use the dedicated cable provided by CoreMorrow to connect P78 and the controller Do not connect via an extension cable casually if you need a longer cable...

Page 13: ...er to achieve better motion precision of P78 it is recommended to use our company s supporting controller Please refer to the user manual of the controller for the specific operation to ensure correct...

Page 14: ...and packing list Check whether the surface of the product has obvious damage Check whether the product identification is clear and complete Check whether the connector is loose and the connector wire...

Page 15: ...Fasten connection Wrong connection Check if the wiring is correct Vibration or inaccurate positioning at startup Incorrect input control parameters Stop immediately and check the parameter Settings Re...

Page 16: ...ount of detergent and then clean it with alcohol Please do not use ultrasonic cleaner When cleaning the surface try not to pick up the product If you have to pick up the product it is better to put th...

Page 17: ...100S K P78 Z200S K P78 Z300S K Unit Active axes Z Z Z Travel range 0 120V 80 150 210 m Travel range 0 150V 100 187 5 262 5 m Unloaded resonant frequency 400 270 200 Hz Load capacity 1 1 0 5 kg Capacit...

Page 18: ...250 CTL as shown below FIG 12 Four core LEMO The sensor connector adopts the four core lemo model LEMO ERA 0s 304 CLL as showne below Note Open loop products have only PZT drive line no sensor line F...

Page 19: ...r model Software driver version Computer operating system that installs supporting software Harbin Core Tomorrow Science Technology Co Ltd Tel 0451 86268790 Fax 0451 86267847 Email info coremorrow com...

Reviews: