E-3
1010FMA-47
Errata
REFERENCE
REFERENCE
QUADRATURE LEADS
FOR NEGATIVE FLOW
QUADRATURE LAGS
FOR POSITIVE FLOW
FOR POSITIVE FLOW
PG3 and PG4 LAGS PG1 and PG2 by 90º
PG1 or PG2
PG3 or PG4
FOR NEGATIVE FLOW
PG3 and PG4 LEADS PG1 and PG2 by 90º
PG1 or PG2
PG3 or PG4
To change the default PGEN settings:
From the [Span Data] menu press
From the [PGEN P/
Unit Volume
] menu cell, press to activate the numeric entry.
Use the numeric keys to enter the desired number of PGEN pulses per unit volume.
NOTE: Ensure that the number entered will provide sufficient resolution while re-
maining within the 20 Hz to 5000 Hz pulse frequency range. Note that the
PGEN output will stop sending pulses at flow rates corresponding to frequen-
cies less than 20 Hz. In this case the PGEN value must be increased.
To store the data press
ENT