2-70
1010PVNFM-3A
Section 2
2.11.4 THE LIQUID DATA MENU
This menu shows the current Liquid Data parameters.
Temp1
Current liquid temperature.
Temp 2
Current liquid temperature.
Reynolds #
Current Reynolds number.
LiquIdent
The LiquIdent™ Index is the sonic velocity of the liquid at 60ºF(15.6ºC).
Pressure
Measured pressure.
Viscosity (cS)
Kinematic viscosity of a liquid measured in centistokes.
LIQUID DATA MENU ITEMS
Temp1
78.5
Temp 2
40.1
Reynolds #
N/A
LiquIdent
1292.80 m/s
Pressure
14.5 PSIA
Viscosity (cS)
1.000 cS
Controlotron Dual Path
SITE 1
RTD 1 temperature sensor reading
Liquid Data
TN uSec
The total elapsed time between the transmission and reception of a sonic
pulse.
TL uSec
The total time a sonic pulse takes to travel through the liquid.
DeltaT nSec
This is the instantaneous result of subtracting the down from the up tran-
sit times.
Note: This will appear to be more active than dampened flow
readings
.
Burst/Sec
Number of transmissions per second under current operating conditions.
The following factors influence this parameter: aeration, pressure trans-
ients, signal strength variation.
% Accepted
The running tally of accepted and rejected up/down burst sets.
Last Makeup
Reason for last Makeup (signal reacquisition).
Makeup Status
Current stage of Makeup routine. [Measurement = normal operation]
APPLICATION INFO MENU ITEMS