Instrument Calibration
44
515 HC04 - 8 June 2007
Time Settings and Data Logging
Instrument Clock
Note:
The real-time clock is part of the advanced option package.
The instrument has a real-time clock for recording logged events. The clock
displays the time and the date. The date format can be set to European
format (day/month/year) or American format (month/day/year). The time
clock uses the 24-hour format.
The clock will continue to operate for up to 5 years (typically) on the
internal battery if there is no power connected to the instrument. Therefore,
after an interruption to the power supply, the instrument recommences
normal operation although there will be no data recorded during the period
without a power supply.
Note:
If there is an interruption to the power supply and the battery has
failed, the instrument displays an error message when the power
supply is restored. In this case, you should set the current time and
date so that the instrument continues to log data at the correct times.
RTU
DATA
The Modbus RTU data format for the 2-register (4-byte) values can be
set as either floating point or long integer values.
Use
or
to select FLOAT or INTEGER.
RTU
ADDR
The Modbus RTU protocol address must be in the range of 1 to 247.
When multiple instruments (slaves) are connected to one communication
device (master), each assigned address must be unique.
Note:
The master device uses the RTU address 0 (zero) for broadcasting
to all connected slave units.
ASCII
ADDR
The ASCII protocol address identifies each communicating device.
The address must be in the range of 1 to 255. When multiple instruments
(slaves) are connected to one computer (master), each assigned address
must be unique.
FLASH
PORT
The Flash Driver Port assignment defines the communication port for
downloading software into the instrument.
The default setting of this assignment is the RS-232 port.
Press
or
to select RS-232, RS-485, or INFRA.
↓
→
UNITS PARAMS INPUTS OUTPUTS ALARMS
COMMS
TM/LOG SETUP TEST END
SET
Summary of Contents for 515
Page 4: ...iv 515 HC04 8 June 2007 ...
Page 8: ...viii 515 HC04 8 June 2007 ...
Page 16: ...Introduction 8 515 HC04 8 June 2007 ...
Page 60: ...Instrument Calibration 52 515 HC04 8 June 2007 ...
Page 80: ...Communications 72 515 HC04 8 June 2007 ...
Page 82: ...74 515 HC04 8 June 2007 ...
Page 88: ...80 515 HC04 8 June 2007 ...