Instrument Calibration
515 HC04 - 8 June 2007
41
The minimum pulse factor required is determined by:
For example: To calculate the required pulse factor to avoid losing counts
in retransmission if a total counts at a maximum rate of 75 units/sec (Hz) and
the required pulse width of a remote counter is at least 50ms:
The maximum pulse output frequency is:
The minimum pulse factor for that frequency is:
Alarms
The alarm relay(s) can be assigned to rate variables such as temperature, or
set as an equipment failure alarm.
The alarm switches “on” whenever an alarm condition exists. The alarm
switches “off” when the alarm condition no longer exists. However, you
may need to configure external alarm devices that require acknowledgement
for cancelling an alarm.
Equipment Failure Alarm
Any alarm relay can be assigned as an equipment failure alarm. This alarm
setting can have normally closed (open) contacts that open (close) when the
instrument displays any error message as listed in
50, or if there is a loss of power to the instrument.
max rate of total
max pulse output frequency
------------------------------------------------------------------
1000
2
50
×
---------------
10Hz
=
75
10
------
7.5Hz
=
↓
→
UNITS PARAMS INPUTS OUTPUTS
ALARMS
COMMS TM/LOG SETUP TEST END
RELAY
ALRM
n
Select a rate variable to assign to the alarm relay.
Note:
If the alarm type is set to “equipment alarm”, this relay assignment
setting is ignored.
Press
or
to select the variable that is required as an alarm.
SET
Summary of Contents for 515
Page 4: ...iv 515 HC04 8 June 2007 ...
Page 8: ...viii 515 HC04 8 June 2007 ...
Page 16: ...Introduction 8 515 HC04 8 June 2007 ...
Page 60: ...Instrument Calibration 52 515 HC04 8 June 2007 ...
Page 80: ...Communications 72 515 HC04 8 June 2007 ...
Page 82: ...74 515 HC04 8 June 2007 ...
Page 88: ...80 515 HC04 8 June 2007 ...