7-8
Maintenance
3 Calibration
Press the wizard’s
Next
button to start the Basic Calibration
process. Color Calibration follows immediately after. The wizard
will inform you when Calibration is complete and instruct you to
remove the Maintenance Sheet.
4 Wrapping
up.
Calibration completes the maintenance process.
Remove the Maintenance Sheet from the scanner. Return the
sheet its protective cover and then place it the storage folder.
Store the folder in a dry place and out of direct light.
7.4 Camera Out of Light Error
The camera out of light error is shown by simultaneous blinking of both
the "Diagnostic" and "Warm Up (Wait)" indicators. The camera out of
light error occurs when one of the CCD cameras moves out of range
and it does not receive enough light to function optimally. This could
happen while making camera adjustments or during transportation of
the scanner.
The
Wait (Warm Up)
indicator will show which camera is out of light by
the how many flashes per period it makes. The camera is indicated by
the following table with A as the leftmost camera and D the rightmost:
Camera A: 1 flash per period.
Camera B: 2 flashes per period.