Setup
—
AI-1204Z-PE Reference Manual
—
23
—
2.
Termination Resistor Setting Switch
The termination resistor setting switch is used to insert the termination resistor. Set up the
termination resistor if suitable for the device to which you are connecting.
This product has internal 50
Ω (±1%)
termination resistors. SW4 sets the resistor for channel 0, SW3
for channel 1, SW2 for channel 2, and SW1 for channel 3.
Do not change the switch setting while the signal source is connected.
Changing the setting may damage this product or connected signal source.
The input current increases when the 50
Ω
termination resistor is inserted.
Check the output current of the input signal source if connecting the termination resistor.
Inappropriate use of the termination resistor may damage the input signal source.
Termination resistor setting
If you want to set a termination resistor, set the switch corresponding to the desired channel to the
connector side.
Termination resistor 50
Ω
setting
[Enabled]
Termination resistor 50
Ω
setting
[Disabled]
<Factory default>
SW1
SW1
SW2
SW2
SW3
SW3
SW4
SW4
3.
Board ID Setting Switch
If you install two or more boards on one personal computer, assign a different ID value to each of
the boards to distinguish them. Set a different value per board.
The board IDs can be set from 0 - Fh to identify up to sixteen boards. If only one board is used, the
original factory setting (Board ID = 0) should be used.
Summary of Contents for AI-1204Z-PE
Page 4: ... 4 Table of Contents Customer Support and Inquiry 82 1 Services 83 ...
Page 19: ... 19 Setup This section explains how to set up this product ...
Page 33: ... 33 Installation This section describes how to connect to an external device with a cable ...
Page 72: ... 72 Appendix This section lists the specifications and the physical dimensions of the product ...
Page 76: ... Appendix AI 1204Z PE Reference Manual 76 2 Physical Dimensions ...
Page 77: ... Appendix AI 1204Z PE Reference Manual 77 3 Circuit Block Diagram ...