Setting up the event controller (Device 2)
The internal clock signal from device 1 is connected to synchronization bus master signal 2
which in turn is connected to synchronization bus slave signal 2 in the synchronization
connector.
Accordingly, setup the event controller to use synchronization bus slave signal 2 as the
sampling clock for device 2.
The signal destination is the sampling clock for analog input and the signal source is
synchronization bus slave signal 2.
9
Start conversion
In this example, conversion is started first on device 2. When analog input starts on device 1,
analog input starts simultaneously on device 2.