Cleaning and Sterilizing
3.1.1
Cleaning Precautions
1.
Follow universal precautions for protective
apparel when handling and cleaning con-
taminated instruments.
2.
Dispose of all non-reusable tubing
sets, irrigation tips, burs, saw blades
and bits properly after use.
3.
Do not lubricate any handpieces.
4.
ConMed Linvatec does not recom-
mend immersing any of our hand-
pieces, as this may affect long-term
reliability.
5.
The use of bleach, chlorine-based,
or sodium hydroxide based liquid or chemi-
cal disinfectants, detergents, enzymatic
cleaners, or soaps may cause degradation of
the aluminum coating on the outside of our
handpieces.
6.
Never clean equipment in an ultrasonic
cleaner or a combination washer/sterilizer.
3.1.2
Handpiece Cleaning Instructions
Clean handpieces and attachments as soon as
possible after use.
1.
Remove all attachments and accessories
(i.e., saw blades, burs) from the handpiece
prior to cleaning. For handpieces with
detachable levers, remove the Activation
Lever (Reference “2.3.4 Removing the
Detachable Handpiece Activation Lever”
on page 16 for removal information).
2.
Disconnect the handpiece cord from the
controller prior to cleaning.
3.
Ensure the handpiece cord is securely fas-
tened to the handpiece during cleaning.
4.
With the nose of the handpiece facing
downward, thoroughly scrub the handpiece,
handpiece cord, attachments, and activa-
tion lever, if applicable, with a clean, soft
brush dampened with a mild, pH-balanced
detergent cleaner of your choice. Remove
all traces of blood, debris and stains.
Remove cleaning solution under running
water. Do Not immerse handpiece.
5.
Manipulate all moving parts of the hand-
piece and attachments to ensure all
debris is removed. If not, continue clean-
ing until all debris is removed.