16
2.3.4
Removing the Detachable
Handpiece Activation Lever
CAUTIONS:
1.
Before attempting to remove the
detachable Handpiece Activation Lever,
ALWAYS place the “SAFE/RUN” Slide
in the “SAFE” position.
2.
Always remove the detachable Activation
Lever when operating the handpiece with
a footswitch.
3.
The handpiece and footswitch operate on
a first come, first served basis. If the
handpiece is activated with the foot-
switch, the handpiece activation lever is
inoperative until the footswitch pedal is
released, and vice versa.
1.
To remove the handpiece detachable
Activation Lever:
(a) Place the “SAFE/RUN” Slide in the
“SAFE” position.
(b) Depress the ribbed portion of the
detachable Activation Lever. Push
towards the “SAFE/RUN” Slide until it
disengages from the handpiece.
2.
To reattach the handpiece Activation Lever:
(a) Insert the Lever pins into the grooves
and reverse the above steps.