14
2.3.2
Preoperative Functional Test for
the MicroPower High Speed Drill
Prior to operating the MicroPower High Speed
Drill, perform the following preoperative tests to
verify proper functioning. Any operating diffi-
culties should be reported to your local Sales
Representative or ConMed Linvatec Customer
Service.
1.
Before operating the drill, check for:
•
any loose or missing parts
•
any physical damage
•
movable parts that do not move freely
2.
Performance Testing:
(a) Assemble the appropriate guard and
bur according to the instructions on
pages 15 through 16.
Warning:
Ensure that the guard and bur are
properly installed, otherwise injury
to the patient or medical personnel
can result.
(b) Operate the drill at 100,000 rpm for
one minute according to operating
instructions contained on pages 15
through 18. Monitor the drill for any of
the following:
•
excessive noise
•
excessive vibration
•
the drill or bur guard are hot to the
touch during the test procedure or dur-
ing surgical use
•
the drill does not operate up to its max-
imum preset speed (verify the maxi-
mum preset operating speed by fully
depressing the drill activation lever or
appropriate footswitch pedal and veri-
fying that the maximum speed is dis-
played on the controller).